7.4 Selecting a program to operate from TASK screen
To select a stored program, press over SEL.
(SEL indication changes its color to indicate it has been selected).
Once selected, this program in placed on the TASK screen.
7.5 Operation screen
On the operation screen «TASK», the program selected on the programming
screen (See 7.4) starts when pressing on
.
This screen shows:
Actual temperature,
Temperature set (set).
Time set (SET).
Elapsed time.
7.6 Datalogging to USB memory
From the “SETTINGS” screen, press the “Download Logs” button to get a text
fi
le with the saved records. If the USB memory stick is not in place, the equi-
pment will wait for the answer.
Fig 7.4 selecting a program
PROGRAM 04
SAVE
TIME: 00:00
TEMP: 50.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
a
SEL
C
a
SEL change to yellow
color when program is
selected
Fig 7.5 Operation screen
HH:MM:SS DD/MM/YY
28.9
ºC
SET
50.0
PROG 04
SET -- h -- m
T
00 h 00m
From the main menu, press over «PROGRAMMING».
Fig7.3 screen is displayed.
Select a program (from 00 to 10) with keys:
Set working temperature and the other parameters:
Press:
keys
will be shown to select temperature.
In a similar way press:
Set de working time.
A time setting value of 00.00 means «continuous
operation».
Sets the low temperature alarm value.
Sets the high temperature alarm value.
Press to save the program.
TEMP: 16.0ºC
UNDERTMP: 0ºC
TIME: 00:00
OVERTMP: 0ºC
SAVE
PROGRAM 00
SAVE
TIME: 00:00
TEMPERATURE: 16.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
Fig 7.3 Programming screen
Selected program no.
Program
selection keys
a
SEL