SYMBOLS

ELECTROMAGNETIC DECLARATION

       WARNING
Use of any accessories and cables other than those specified in the manufacturer’s documentation, with the exception of cables sold by the manufacturer as 
replacement parts for internal components, may result in increased emissions or decreased immunity of the Icare TONOVET Plus tonometer.

       WARNING
Use of any accessory or cable with the Icare TONOVET Plus tonometer other than those specified may result in increased emissions or decreased immunity of the 
Icare TONOVET Plus tonometer.

SN

LOT

Attention!!! See instructions

See operating instructions  
for more information

Serial number

Single use only

BF-type device

Lot number

Manufacturing date

Keep dry

Manufacturer

Storage environment

Transport environment

PERIODIC SAFETY CHECKS

We recommend that the following checks be performed every 24 months.
Equipment inspection for mechanical and functional damage.
Inspection of safety labels for legibility.

MAINTENANCE

Follow local regulations and recycling instructions regarding the disposal or recycling of the Icare TONOVET tonometer and accessories.

 

WARNING

The tonometer should only be opened by qualified service personnel. It contains no userserviceable parts, apart from the batteries and a probe base.  
The Icare TONOVET tonometer requires no routine servicing or calibration other than changing the batteries at least every 12 months and changing or cleaning  
the probe base. If servicing is necessary, contact your local Icare representative.

REPLACING/CLEANING THE PROBE BASE

During the use some dirt may collect in the probe base, affecting the probe movement. Replace the probe base every 12 months. Clean the probe base 
every 6 months. Clean or replace the probe base if the error message Clean Change is displayed. 

Instructions for replacing the probe base: 

• Replace every 12 months.
• Turn off the tonometer. 
• Unscrew the probe base collar.
• Remove the probe base by tilting the tonometer downwards  
  and pulling the probe base out of the tonometer. 
• Insert a new probe base into the tonometer. 
• Screw the collar in, to lock the probe base.

Instructions for cleaning the probe base

• Clean every 6 months. 
• Fill the probe base cleaning container or other clean container with 70-100 % isopropyl alcohol. 
• Turn the power off. 
• Unscrew the probe base collar.
• Remove the probe base by tilting the tonometer downwards  
  and pulling the probe base out of the tonometer. 
• Insert the probe base into the cleaning container and let soak for 5–30 minutes. 
• Remove the probe base from alcohol. 
• Dry the probe base by blowing clean canned or compressed air into the hole  

 

 

  in the probe base. This will additionally remove possible residual dirt. 
• Insert the probe base into the tonometer. 
• Screw the collar in, to lock the probe base.

CLEANING THE TONOMETER

    

WARNING

Never immerse the Icare TONOVET tonometer, spray, pour or spill liquid onto the Icare TONOVET tonometer, its accessories, connectors, switches or open-
ings in the chassis. Dry any liquid on the surface of the tonometer immediately. 

Icare TONOVET tonometer’s surfaces have been tested and found chemically resistant to the following liquids: 
• 70-100 % isopropyl alcohol
• Mild soap solution 
• 95 % Pursept solution 

Cleaning instructions for surfaces: 
• Turn the power off. 
• Dampen a soft cloth with one of the liquids mentioned above. 
• Lightly wipe the surfaces of the tonometer with the soft cloth. 
• Dry the surfaces with a dry soft cloth.

RETURNING THE ICARE TONOVET TONOMETER FOR SERVICING / REPAIR

Contact your local Icare representative for shipping instructions. Unless otherwise instructed by Icare Finland, there is no need to ship accessories along 
with the tonometer. Use a suitable carton with the appropriate packaging material to protect the device during shipment. Return the device using any 
shipping method that includes proof of delivery.

Do not discard this product with other household-type waste. Send to appropriate
facility for recovery and recycling. EU WEEE (European Union Directive for Waste of
Electronic and Electrical Equipment)

Temperature limits 

Humidity limits 

Atmospheric pressure limits 

-10°C

55°C

95%

10%

1060hPa

700hPa

-40°C

70°C

95%

10%

1060hPa

500hPa

Summary of Contents for TONOVET PLUS

Page 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Page 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Page 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Page 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Page 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Page 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Page 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Page 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Page 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Reviews: