Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

Icare 

TONOVET

 Plus tonometer instruction manual

ENGLISH

www.tonovet.com

4

www.tonovet.com

5

TURNING THE TONOMETER ON

Place the wrist strap around your wrist and secure it. The wrist strap protects the tonometer from dropping onto the floor accidentally.

To turn the tonometer on press the Select or Measure button. The following are illustrations of these two alternative ways of starting the tonometer:

CHANGING THE SETTING FOR DIFFERENT SPECIES

To change the setting for different species press the Select button to access the menu. Press Navigation buttons and select SETTING by pressing the Select button. 
Toggle the settings by Navigation buttons and select a specie setting by pressing Select button. To measure return to MEASURE by navigation buttons.

PROBE BASE LIGHT INDICATION

The probe base light serves two purposes. First, it helps guide alignment of the probe by showing a red light when the device is in the wrong position (i.e. too 
much vertical tilt) and a green light when the orientation is correct. Second, it indicates errors (see section Error and Info Messages) in addition to the display 
during the measurement sequence. When any of these errors occurs, the probe base light flashes red until the error is cleared by pressing the Measure button.  
The probe base light flashes red also when the yellow or red lights are displyed indicating a measurement sequence with high deviation.

INTENDED USE

The Icare TONOVET Plus tonometer is used in the intraocular pressure (IOP) measuring in the veterinary medicine. The Icare TONOVET Plus tonometer is intend-
ed for veterinary use only.

INTRODUCTION

The Icare TONOVET Plus tonometer is based on a patented, induction-based rebound method, which allows intraocular pressure (IOP) to be measured accurately, 
rapidly and without anesthetic.

The tonometer uses the rebound method. A small and light single-use probe makes contact with the eye very briefly. The tonometer measures the deceleration 
of the probe and the rebound time, and calculates the IOP from these parameters. 

A measurement sequence includes six measurements. The probe moves to the cornea and back during every measurement. As a result, after the six measure-
ments the tonometer calculates the final IOP and stores it in the tonometer’s memory. 

IOP changes due to the effects of pulse, breathing, eye movements, and body position. Because the measurement is made handheld in fractions of a second, six 
measurements are needed to obtain the final reading.

Icare TONOVET Plus tonometer has settings for dog, cat, rabbit and horse.

 

PACKAGE CONTENTS

   NOTE!
When you have opened the package, check for any external damage or flaws, particularly for damage to the case. If you suspect that there is something wrong 
with the tonometer, contact the manufacturer or distributor.

The package contains:

• 

Icare TONOVET Plus tonometer

• 

4 x AA batteries

• 

100 single use probes in a box

• 

aluminum case

• 

IOP pad

• 

probe base cleaning container

• 

probe base collar, narrow 

• 

probe base cover

• 

quick guide

• 

screw driver 

• 

silicone grip

• 

spare probe base

• 

USB memory stick including instruction manuals (as PDF)

• 

warranty card

• 

wrist strap

PARTS OF THE TONOMETER

1.

 Measuring support 

2.

 Probe base

3.

 Collar

4.

 Display

5. 

Measuring support adjusting wheel

6.

 Navigation buttons

7.

 Measure button

8.

 Select button

1. 

5.

4.

3.

2.

6.

8.

LOADING THE PROBE

Remove the red probe base  
cover (do not dispose of the cover) 
and open the probe tube by  
removing the cap and insert the 
probe into the probe base as  
shown in the image. 

After loading the probe the 
tonometer will be ready for 
measurement when Play-symbol and 
active specie setting appears on the 
display.

7.

Battery compartment cover

INSTALLING OR CHANGING THE BATTERIES 

Unscrew the battery compartment locking screw with a screwdriver. Remove the battery compartment cover. 
Place the wrist strap into the wrist strap attachment at the end of the tonometer. 

Insert a new set of four AA batteries. Insert the batteries accordance with the picture below, installation order 1 to 4. Do not use rechargeable batteries.

Replace the battery compartment 
cover and secure it by screwing 
it in lightly using the screwdriver.
Take care not to use excessive 
force when screwing the 
cover into place.

Pressing the Select button

Pressing the Measure button

1.

2.

3.

4.

Summary of Contents for TONOVET PLUS

Page 1: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 1 ENGLISH...

Page 2: ...er contact the manufacturer or distributor Use the tonometer only for measuring intraocular pressure Any other use is improper and the manufacturer cannot be held liable for any damage arising from im...

Page 3: ...he IOP from these parameters A measurement sequence includes six measurements The probe moves to the cornea and back during every measurement As a result after the six measure ments the tonometer calc...

Page 4: ...ull or yellow some variation segments If variation is too big red segments will be displayed Yellow Repeat symbol will be dispayed in case of external distubance eg EMC during the measurement The disp...

Page 5: ...he Menu functions starts from the MEASURE display press either of the Navigation buttons located around the Select button Menu functions are MEASURE SETTING HISTORY SOUND LIGHT LANGUAGE and INFO MEASU...

Page 6: ...IMENSIONS 104 Box of 100 probes 89 g 53 x 109 x 36 mm 7217 Probe base collar 4 g 18 x 18 mm 7218 Probe base collar narrow 1 g 17 x 18 mm 540 Probe base 4 g 7 x 38 mm 543 Probe base cleaning container...

Page 7: ...Remove the probe base by tilting the tonometer downwards and pulling the probe base out of the tonometer Insert a new probe base into the tonometer Screw the collar in to lock the probe base Instructi...

Page 8: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 NOT APPLICABL...

Page 9: ...Icare TONOVET Plus tonometer instruction manual ENGLISH www tonovet com 16 NOTES...

Reviews: