IC Sciences TMB-1018-BT User Manual Download Page 8

13

12

DATA MANAGEMENT

DATA MANAGEMENT

2

the records. Up to 60 records will be stored

under each user ID.

The measurement date and time will be

displayed alternatively.

Recall the Records

2

.Press MEM button and SET button to rotate

1.When the monitor is OFF, press MEM

button to retrieve the memory. The monitor 

will display the average value of latest three 

measurements. 

3

.If you would like to check another user

ID’s history, please follow the

instructions in Select User ID to

change to another User ID first.

Delete the Records

1.When under data enquiry mode, press

and hold both MEM button and SET button 

for 3 seconds to clear memory. The LCD 

will display “dEL dOnE”, indicating that the 

memory is cleared.

No Record found for User 2!

Transmitting Data

Manual Data Transmission

With the advanced Bluetooth 4.0 technology applied, the mobile or portable 

equipments, which are equipped with Bluetooth function in line with BLE 

Technical Specifications as well as BLP Protocol established by global

organization Bluetooth SIG, are capable to receive your personal health data.
When both the My BP Coach app and Bluetooth are ON, TMB-1018-BT

 

 

1. After measurement, the symbol       lights

  up, indicating the measuring result is being 

  automatically transmitted to smartphone.

2

.If successful, the LCD will display

“dOnE”.

3

.If the devices fail to connect, the LCD will

display “ERROR”.

CAUTION

Interference may occur in the vicinity of equipment marked with the following symbol

      . And TMB-1018-BT may interfere other electrical equipment in the area.
Sensitive people, including pregnant women and those with implanted medical elec- 

tronic devices, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters (about 8 inches) away from the human 

body (especially the head) when the data transmission is proceeding after 

measurement.
To enable the data transmission function, this product should be paired to Bluetooth

end at 2.4 GHz.

How to mitigate  possible interference?

1. The range between the device and BT end should be reasonably close, from 1

meter to 10 meters (about 3 to 30 feet). Please ensure no obstacles are between 

the device and BT end to obtain a connection and to lower the RF output range.

2. To avoid interference, other electronic devices (particularly those with wireless

transmission/Transmitter) should be kept at least 1 meter (about 3 feet) away 

from the monitor.

will automatically transmit measurement data to your smartphone via Bluetooth.

Summary of Contents for TMB-1018-BT

Page 1: ...read the user manual carefully and thoroughly to ensure safe usage of this product and keep the manual available for future reference Thank you for selecting the IC Sciences Blood Pressure Arm Cuff Monitor TMB 1018 BT Contains FCC ID OU9AW8001 LS GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO LTD Zone A 5 F Investment Building No 12 Huizhan East Rd Torch Development District Zhongshan Guangdong 528437 ...

Page 2: ...ta INFORMATION FOR USER 14 Tips for Measurement Maintenance START MEASUREMENT 11 Start Measurement TABLE OF CONTENTS ABOUT BLOOD PRESSURE 16 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why are measurements taken in hospital different from at home Is the result the same when measuring on...

Page 3: ...re taking medication consult your physician to determine the most appropriate time for your measurement Never change a prescribed medication without your physician s consent This unit is not suitable for continuous monitoring during medical emergencies or operations If the pressure of the cuff exceeds 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when its pressure...

Page 4: ...ood pressure Error Memory Review Mode Average Value Current Time Month Day Hour Minute Irregular Heartbeat Detector Indicate the current mode is memory review mode and which group of memory it is Bluetooth is turned on Detects irregular heart beats Average value of latest three records The monitor detects error AIR CONNECTOR PLUG List 1 Blood Pressure Monitor TMB 1018 BT 4 User manual 3 4 AAA batt...

Page 5: ... cuff s air pipe into the interface located on the right side of the monitor 3 Tighten the cuff Make sure the cuff is fixed 2 to 3 centimeters about an inch above your elbow You should be able to insert one finger when the cuff is tightened around your upper arm 4 Correct Posture It is best to position the cuff on your bare arm Sit comfortably and relaxed in a chair The center of the cuff should b...

Page 6: ...ess and hold SET button to enter HOUR and MINUTE setting 2 Press MEM button to change the numeral Each press will increase the numeral by one in a cycling manner 3 Press SET button to confirm the HOUR and MINUTE Then the monitor diverts to MONTH and DAY setting YEAR 6 After confirming the YEAR the LCD will display dOnE and the monitor will shut off Select User ID NOTE The monitor will shut off 60 ...

Page 7: ...Device 1 Turn on Bluetooth and the My BP Coach app Make sure both are ON when pair up is proceeding 2 With the monitor OFF press and hold the START STOP button about five seconds to start pair up until the symbol blinks indicating pair up is proceeding If successful symbols will be shown on the LCD like the picture on the right If devices fail to connect symbols will be shown on the LCD like the p...

Page 8: ...B 1018 BT 1 After measurement the symbol lights up indicating the measuring result is being automatically transmitted to smartphone 2 If successful the LCD will display dOnE 3 If the devices fail to connect the LCD will display ERROR CAUTION Interference may occur in the vicinity of equipment marked with the following symbol And TMB 1018 BT may interfere other electrical equipment in the area Sens...

Page 9: ...unctions safely and is in proper working order before use Don t service or maintain while the device is in use 2 If you have any problems with this device such as setting up maintaining or using please contact IC Sciences Do not open or repair the device yourself 3 Please contact IC Sciences if any unexpected operation or events occur 4 Cleaning Dusty environments may affect the performance of the...

Page 10: ...elax the blood pressure reaches its minimum value in the cycle which is called diastolic pressure Irregular Heartbeat Detector This Blood Pressure Monitor is equipped with an intelligent function of Irregular Heartbeat IHB Detector During each measurement the arm cuff records the heartbeat intervals and determines the standard deviation If the calculated value is larger than or equal to 15 this mo...

Page 11: ...r a moment and then measure again The measurement process does not detect the pulse signal Loosen clothing on the arm and then measure again Relax for a moment and then measure again The treatment of the measurement failed Power supply Battery Powered Mode 6V 4 x AAA size batteries AC Adaptor Powered Mode 6V 1A Can be supplied by AC adaptor model UE08WCP 060100SPA Included Display mode Blue LCD wi...

Page 12: ...not occur in a particular installation If the product does cause harmful interference to radio or television reception which can be determined by turning the product on off the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and the receiver Connect the product into...

Page 13: ...and MANUFACTURER S declaration electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS Table 4 Guidance and MANUFACTURER S declaration electromagnetic IMMUNITY for ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING ...

Page 14: ...24 25 Table 6 Recommended separation distances between portable and mobile RF communications equipment and the ME EQUIPMENT or ME SYSTEM for ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING ...

Reviews: