background image

 

3

 

 

1.

 

Installation Instructions 

 

1.

 

Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface. 

2.

 

Ensure  a  good  working  environment  free  of  any  combustible  media  or  flammable  materials  to  avoid 
additional hazards. 

3.

 

Place  the  shaker  at  a  distance  of more  than  10cm  from  the wall  and  other  units  if  using  multiple units 
together. 

4.

 

Ensure that there is no accidental spillage during speed setting. 

5.

 

Do not use any combustible media or flammable materials to avoid any additional hazards. 

6.

 

Ensure that the label indicates the correct voltage before connecting the unit to the power supply. 

7.

 

Do not operate the unit with a damaged power cord. 

8.

 

Disable the power supply while fitting the accessories. 

9.

 

Ensure that the unit and its accessories are in good working condition before each operation. 

 
 
  

2.

 

Operating Procedure 

 

1.

 

Switch ON the power and the instrument will start self-checking. 

2.

 

Press the “Mode” key and the unit will enter the adjustment mode

3.

 

Press  the  “+/

-

”  keys  and  confirm  the  set  speed  and  time.  Adjust  the  speed  and  time  to  the  desired 

settings. 

4.

 

Press the “Start” key and

 the instrument will begin shaking. 

5.

 

Press the “Stop” key to stop the operation and return the unit to adjustment

 mode. 

 
 
 

3.

 

Troubleshooting and Error Code Information 

 

Error Code 

Problem 

Reason 

Solution 

E01 

No operating 

response found (LED 

OFF) 

Power switch is OFF 

Check  and  connect  the  power  supply 
before restarting. Check the adapter. 

Switch position is OFF 

Turn  on  the  instrument  and  check  the 
speed set on the LED display. 

Internal cables disconnected 

Open  the  instrument  and  check  all 
connections before reconnecting them. 

E02 

The shaking board 

operation failure 

Target speed is not set 

Set  the  target  speed  and  make  sure 
the indicator lamp is ON. 

Driver board damage 

Replace the driver board. 

Motor damage 

Replace the motor. 

E03 

Inaccurate shaking 

speed 

Driver board damage 

Replace the driver board. 

Photoelectric switch damage  Replace the photoelectric switch. 

E04 

Shaking speed is 

jammed 

Driver board damage 

Replace the driver board. 

Motor damage 

Replace the motor. 

 
 
 
 
 
 

Reviews: