3
1.
Installation Instructions
1.
Carefully unpack the instrument and place it on a stable, clean, non-slip, and fireproof surface.
2.
Ensure a good working environment free of any combustible media or flammable materials to avoid
additional hazards.
3.
Place the shaker at a distance of more than 10cm from the wall and other units if using multiple units
together.
4.
Ensure that there is no accidental spillage during speed setting.
5.
Do not use any combustible media or flammable materials to avoid any additional hazards.
6.
Ensure that the label indicates the correct voltage before connecting the unit to the power supply.
7.
Do not operate the unit with a damaged power cord.
8.
Disable the power supply while fitting the accessories.
9.
Ensure that the unit and its accessories are in good working condition before each operation.
2.
Operating Procedure
1.
Switch ON the power and the instrument will start self-checking.
2.
Press the “Mode” key and the unit will enter the adjustment mode
.
3.
Press the “+/
-
” keys and confirm the set speed and time. Adjust the speed and time to the desired
settings.
4.
Press the “Start” key and
the instrument will begin shaking.
5.
Press the “Stop” key to stop the operation and return the unit to adjustment
mode.
3.
Troubleshooting and Error Code Information
Error Code
Problem
Reason
Solution
E01
No operating
response found (LED
OFF)
Power switch is OFF
Check and connect the power supply
before restarting. Check the adapter.
Switch position is OFF
Turn on the instrument and check the
speed set on the LED display.
Internal cables disconnected
Open the instrument and check all
connections before reconnecting them.
E02
The shaking board
operation failure
Target speed is not set
Set the target speed and make sure
the indicator lamp is ON.
Driver board damage
Replace the driver board.
Motor damage
Replace the motor.
E03
Inaccurate shaking
speed
Driver board damage
Replace the driver board.
Photoelectric switch damage Replace the photoelectric switch.
E04
Shaking speed is
jammed
Driver board damage
Replace the driver board.
Motor damage
Replace the motor.