Biotech-Lab User Manual
\\
24
www.ibpmedical.com
7.5.1 PH SENSOR ADJUSTMENT
Return
Value 1 adjustment
Allows to select the number of adjustment points.
Current reading of pH and temperature.
Allows to select and input the buffer solution value.
Leads to the next calibration point.
Return
Value 2 adjustment
Allows selecting the number of adjustment points.
Current reading of pH and temperature.
Allows selecting and inputting the buffer solution value.
Leads to the next calibration point if any.
Otherwise the next window shows the calibration result.
Once all adjustment points are finished the adjustment values are store automatically in the sensor. If something is
wrong with the adjustment the sensor will not accept the values and an error message will be shown.
The Sensor calculates slope of the pH-Probe. It should be better than 75%. Below 75% the pH-Electrode is
considered defective.