ibidi Gas Incubation System – Silver Line
1. Place the Gas Mixer and Humidifying Column onto the microscope table next to the micro-
scope.
2. Confirm that the power switch of the Gas Mixer is off (power switch is on the back of the Gas
Mixer).
3. Connect the gas tubing from the gas supply to the back of the Gas Mixer (Section
4. Connect the Heated Tubing Package (see Section
5. Connect the humidity sensor cable, and the power supply (Section
and
Figure
shows all connections of cables and tubing on the back of the Gas Mixer.
5A
Heating
Humidifying Column
Bridge cable from
Temperature Controller
USB connection (optional)
External power supply
Type:
Serial number:
Production date:
ibidi Gas Mixer
ibiGM2-001
11/2018
ibidi GmbH
Am Klopferspitz 19
D-82152 Martinsried
Germany
+49(0)89520 46 17 0
www.ibidi.com
Output 1
Input N
2
Input Air
Input CO
2
Output 2
Figure 10: Overview of the connection of all cables and tubing to the Gas Mixer. In the CO
2
version
the N
2
port is not available as closed by a seal.
5.1.1
Connection to the Gas Supply
The gases from the bottle or laboratory gas line need to be connected to the Gas Mixer with the OD 6
mm polyurethane tubing of the respective color. This tubing is delivered with the system.
Table 5: Available connections and color code in the two different Gas Mixer versions
Gas Input
Gas Pressure
Tubing Color
Air
1 bar/14.5 psi
blue
CO
2
1 bar/14.5 psi
green
N
2
*
1 bar/14.5 psi
yellow
*CO
2
/O
2
version only
Version 1.0 (June 1, 2022)
24
Instruction Manual