Push
Figure 14: How to connect the gas tubing to a pressure reducing valve from a gas line, compressor,
or gas bottle. A push-in fitting for 6 mm O.D. tubing is recommended.
WARNING
– Only disconnect the tubing when there is no pressure on the gas line. First,
switch off the pressure from the gas line. Control the remaining pressure on the gas line’s pressure
reducing valve. Then, remove the tubing.
5.1.2
Connection of the Heated Tubing Package and the Humidifying Column
There are two lines coming from the Gas Mixer output. One goes to the Humidifying Column, the
other directly to the Incubation Chamber.
Table 6: Connection of the gas lines with color code of the mixed gas
Color Code
Gas Mixer output 1 (white)
Blue tubing with white label
Gas Mixer output 2 (yellow)
Blue tubing with yellow label
Humidifying Column output red
Blue tubing with red label
Humidifying Column output blue
Blue tubing with blue label
1. Connect the two gas outlets of the Gas Mixer to the white (1) and yellow (2) marked gas tubing
to Output 1 and Output 2 on the back wall of the Gas Mixer.
2. Connect the plug of the electrical cable (next to the white and yellow gas tubing) to the ”Heater”
socket.
3. Connect the blue and red marked gas tubing to the respective inputs on the Humidifying Col-
umn.