OPERATION OF INSTRUMENT 

Power on: 

Press and hold the power button until the 
LCD shows the PDVS2mini splash screen. 

Power off: 

Press and immediately release the power 
button. The LCD will clear, power is now off. 

 

The unit will provide a low battery warning at 
13.0Vdc and will auto shutdown at 12.5Vdc. 
Both settings are adjustable via settings. 

Initialize: 

The splash screen is displayed on power up with the software version before 
displaying the MAIN screen. It is recommended to wait 10mins from cold to allow the 
LM399AH reference to stabilize. 

Menu: 

To access the MENU, press and hold the ENTER button. 

 

The Main Menu functions are as follows, 
 

UP 

– MAIN 

 

 

The main output control screen 

 
LEFT 

– SETTINGS   

Where user config settings can be changed. 

 
RIGHT 

– CALIBRATION 

Where calibration of the unit can be done. 

Power source: 

The primary source of power is the batteries. The best performance of the unit is 
when the batteries are being used to supply power without any external power. 

External DC input cable wiring: Red +, Blk -. 
 
The battery terminals and the external DC input are reverse polarity protected. 
 
Use only matched charged state 

batteries, i.e. don’t fit two batteries that are not 

charged to the same capacity/level. 

The batteries must be capable of charging at a rate of 220mA. Most are, but please 
check. 
 
The batteries are charged in-place using the DC socket on the top. The PDVS2mini 
must be powered up for charging to take place. This is so the PDVS2mini can 
actively monitor the charging process. 

The batteries won’t charge with the 

PDVS2mini turned off. 
 
The unit can be used whilst being powered from the ext. DC socket but with the 
following important restrictions/notes: 

 
Minimum DC supply (to charge the batteries): 

= 18Vdc 

Minimum DC supply (no batteries fitted): 

 

= 18Vdc 

Maximum DC supply:   

 

 

 

= 24Vdc 

 
Absolute Minimum DC supply: 

= 14Vdc 

(unit won’t charge but will run) 

Absolute Maximum DC supply: 

= 24Vdc 

 

 

Summary of Contents for PDVS2mini

Page 1: ...Model PDVS2mini A multitude of uses as a calibrator reference and fully adjustable precision voltage source Handmade in Great Britain by Ian Johnston Manual Version 1 0 19 10 19 Ian Johnston Engineer...

Page 2: ...FIRMWARE UPGRADE 9 10 WINDOWS PC APPLICATION 11 HARDWARE 12 FREQUENTLY ASKED QUESTIONS 13 PRODUCT DISCLAIMER WARRANTY 14 CE CERTIFICATE DECLARATION 15 CALIBRATION RECORD BLANK 16 The screenshots phot...

Page 3: ...oltage reference Uses an LM399AH 0 5 1 ppm degC temperature coefficient 0 00000 Vdc to 10 00000 Vdc in 10 uV 0 00001 Vdc steps Load 3kohm minimum Example At 10Vdc output with 3kohm load 3 3mA whilst m...

Page 4: ...is the batteries The best performance of the unit is when the batteries are being used to supply power without any external power External DC input cable wiring Red Blk The battery terminals and the e...

Page 5: ...n off charging via the SETTINGS menu Without this you may get OVER on the display Please observe the minimum external DC supply voltage NOTES When running with batteries and also an external DC input...

Page 6: ...l Best Stability For the most accurate and stable output it is highly recommended to run the unit from the batteries with no external DC power input This helps avoids any ground loop noise entering th...

Page 7: ...col BAUD 250k N 8 1 Rate Data from PDVS2mini to PC is sent every 100mS Format to PC VR0 0 12345 Format from PC KeyVoltage 0 1 23456 The following details all the data being sent from the PDVS2mini and...

Page 8: ...imum current should be approx 250mA so the default setting here of 400mA gives some margin Charge Full mA Current mA at which the battery is determined to be near full The charging circuit is self con...

Page 9: ...10 00000 XXXXXXX Counts for 10 00000Vdc output Out Vdc XX X XXXX Output monitor counts Calibration of the output DC voltage monitor Batt DC Vdc XX X XXXX Battery Ext power supply Vdc Span counts Cali...

Page 10: ...le co uk avrdudess a gui for avrdude A HEX file will be supplied for the upgrade The current calibration settings will not be overwritten but it is advisable to have them written down Please refer to...

Page 11: ...oth ends type from the PDVS2mini ICSP header to the AVR Dragon ISP header note orientation and pin1 On Atmel Studio Select AVR DRAGON ATmega1284 ISP then hit APPLY The PDVS2mini will be interrogated y...

Page 12: ...PC then double click on SETUP EXE Follow the on screen prompts to install the app If the app install fails then see below Turn on COMMS on the PDVS2mini by holding down the LEFT RIGHT buttons Start th...

Page 13: ...e Pcb printed circuit board FUSE 1 1A time delay fuse Schurter UMZ 250 series J1 3 3Vdc supply to ICSP header Default shorted ICSP In Circuit Serial Programming header used for uploading new firmware...

Page 14: ...d charge them separately in a standalone charger This will match them again and you can return them to the PDVS2mini If this problem persists then you may have a faulty weak battery Q Can I run the PD...

Page 15: ...product may directly threaten life or injury The user of Ian Johnston Engineering Ltd products in life support aircraft applications or other such critical activities assumes all risks of such use an...

Page 16: ...15...

Page 17: ...00 4 00000 5 00000 6 00000 7 00000 8 00000 9 00000 10 00000 FUNCTION COUNTS Out Vdc Batt Vdc Charge mA Batt Low Shutdown Charge Enable Charge Overload mA Charge Full mA TEST Serial Comms Soak Test Dur...

Reviews: