3
OPERATION OF INSTRUMENT
Power on:
Press and hold the power button until the
LCD shows the PDVS2mini splash screen.
Power off:
Press and immediately release the power
button. The LCD will clear, power is now off.
The unit will provide a low battery warning at
13.0Vdc and will auto shutdown at 12.5Vdc.
Both settings are adjustable via settings.
Initialize:
The splash screen is displayed on power up with the software version before
displaying the MAIN screen. It is recommended to wait 10mins from cold to allow the
LM399AH reference to stabilize.
Menu:
To access the MENU, press and hold the ENTER button.
The Main Menu functions are as follows,
UP
– MAIN
The main output control screen
LEFT
– SETTINGS
Where user config settings can be changed.
RIGHT
– CALIBRATION
Where calibration of the unit can be done.
Power source:
The primary source of power is the batteries. The best performance of the unit is
when the batteries are being used to supply power without any external power.
External DC input cable wiring: Red +, Blk -.
The battery terminals and the external DC input are reverse polarity protected.
Use only matched charged state
batteries, i.e. don’t fit two batteries that are not
charged to the same capacity/level.
The batteries must be capable of charging at a rate of 220mA. Most are, but please
check.
The batteries are charged in-place using the DC socket on the top. The PDVS2mini
must be powered up for charging to take place. This is so the PDVS2mini can
actively monitor the charging process.
The batteries won’t charge with the
PDVS2mini turned off.
The unit can be used whilst being powered from the ext. DC socket but with the
following important restrictions/notes:
Minimum DC supply (to charge the batteries):
= 18Vdc
Minimum DC supply (no batteries fitted):
= 18Vdc
Maximum DC supply:
= 24Vdc
Absolute Minimum DC supply:
= 14Vdc
(unit won’t charge but will run)
Absolute Maximum DC supply:
= 24Vdc