I-Tech WEPERE RESTART User Manual Download Page 15

15   

USER MANUAL

The device does not require sterilisation.

Notes:

-   Never use solvents for cleaning. Cleaning agents can damage the device.
-   Carry out routine maintenance, in particular:
   •  inspect the body of the device for cracks or fissures, which may allow liquids to enter;
  • inspect the cables.

TRANSPORT AND STORAGE
Transport precautions

There is no particular care to be taken during transport as RESTART is a portable device. However, 

it is recommended to put RESTART and its accessories in the bag provided after each use. Protect 
the device from intense heat, direct sunlight and liquids. Store the device in a cool and well venti-
lated environment. Do not place heavy objects on top of the device.

Storage precautions

The storage location should have the following characteristics:

During operation

 

ambient temperature from +5 to + 40°C

   

relative humidity from 15 to 93%

   

pressure from 700 to 1060 hPa

In the bag provided

 

ambient temperature from +5 to +40 °C

   

relative humidity from 15 to 93%

   

pressure from 700 to 1060 hPa

Troubleshooting

Any type of work on RESTART must only be carried out by the manufacturer or authorised dealer. 
In any case, before sending RESTART to the manufacturer, it will be necessary to ascertain the 
exact nature of the RESTART malfunction. Check the following:

PROBLEM

POSSIBLE CAUSE

SOLUTION

The device does not turn on

Adapter contact error

Make sure the adapter 

is connected.

Check the following contacts:

• All contacts are fine

• All contacts are not broken

• The battery is O

The device does not work

The battery is damaged

The LED indicators do not light up

Mains plug not inserted correctly 

in the power socket.

Check the operation of the power 

socket.

Mains cable not correctly 

inserted in the connector 

of the device.

Insert the plug and cable 

correctly into the connector 

of the appliance.

Mains cable worn and broken.

Replace the mains cable.

Summary of Contents for WEPERE RESTART

Page 1: ...USER MANUAL Recover faster anytime anywhere...

Page 2: ......

Page 3: ...e description and controls 7 Labelling 8 Pack contents 10 How To Use Introduction to the technology 10 Contraindications 11 Warnings 12 Device use 12 Treatments 14 Looking after the device Maintenance...

Page 4: ...ns of the electromagnetic compatibility Directive 2014 30 UE of the European Parliament and of the Council of the 26th February 2014 to the current TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY...

Page 5: ...able anaesthetic mixture with air or with oxygen or with nitrous oxide device intended for continuous operation device not suitable for external use Intended purpose and scope of use The RESTART devic...

Page 6: ...0Hz 10 Duty cycle 5 50 100 mains powered 5 50 100 battery powered Power Can be set in 3 steps L low M medium H high Maximum power density 1 6W cm mains powered 0 8W cm battery powered Maximum output p...

Page 7: ...L Device description and controls 1 Treatment time indicator 2 Time button 3 Power adjustment button 4 Treatment power indicator 5 Battery indicator 6 Device ON OFF button 7 Emitting head 8 Power supp...

Page 8: ...Out 15V 1 2A Battery Ni MH AAA850mAh 4 8 V Ultrasound Carrier frequency of use 1MHz 10 Duty cicle 5 50 100 mains powered 5 50 100 battery powered Maximum output power 6 4W mains powered 3 2W batter po...

Page 9: ...documents accompanying the product IP22 Device protected against the penetration of solids with diameter d 12 5mm and against the fall of vertical drops of water when the device is held at 15 from the...

Page 10: ...sound waves would have to over come to reach the body To reduce this difference a coupling substance must be used If there is a small amount of air between the transducer and the skin the proportion o...

Page 11: ...advised not to wear metal objects to use ONLY the accessories supplied by the manufacturer It is forbidden to use the device in the presence of equipment for monitoring the user s vital functions equi...

Page 12: ...s used in strict compliance with the instructions given in this manual Should any foreign substances get into the device contact the dealer or manufacturer immedia tely Should the device fall check th...

Page 13: ...has been set and the head has been placed in contact with the skin the treatment will begin it is important to perform the treatment by continuously and uniformly moving the head around the ap plicat...

Page 14: ...manufacturer Treatment Intensity Minutes Muscle H Power supply Battery 15 Tendon Ligament L M Power supply M Battery 10 15 Aesthetic L Power supply L M Battery 5 10 REMEMBER TO keep the ultrasound hea...

Page 15: ...elative humidity from 15 to 93 pressure from 700 to 1060 hPa In the bag provided ambient temperature from 5 to 40 C relative humidity from 15 to 93 pressure from 700 to 1060 hPa Troubleshooting Any ty...

Page 16: ...ry indicator will flash from L M H in sequence Once char ging is complete the battery charge indicator will light up as H High WARNING the battery life cycle depends on the charge discharge cycles to...

Page 17: ...ided by IACER Shouldyouneedtoreturnthegoodsthenpleasepackthedeviceandalltheaccessoriessothatitwon t be damaged during transportation In order to be entitled to the warranty assistance the purchaser mu...

Page 18: ...facturer if non original spare parts are used the operation and safety of the product might be affected and the warranty will be null and void Interference and electromagnetic compatibility tables The...

Page 19: ...ns and to use caution in therapeutic applications on users with pacemakers However it is advisable to use the device keeping a distance of at least 3 metres from televisions monitors mobile phones or...

Page 20: ...USER MANUAL 20 RESTART All rights reserved RESTART and the logo are the exclusive property of I A C E R S r l and are registered trademarks Issue MNPG417 01 of 31 01 2022...

Page 21: ...21 USER MANUAL...

Page 22: ...omagnetic environment guidance RF emissions CISPR 11 Group 1 The RESTART uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interfere...

Page 23: ...line 2kV Line s to line 1kV Line s to line 2kV Line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variation...

Page 24: ...nsmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compl...

Page 25: ...ation distance according to the frequency of the transmitter m d 1 2 P from 150kHz to 80MHz d 1 2 P from 80MHz to 800MHz d 2 3 P from 800MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 01 0 38 0 38 0 73 1 1 2 1...

Page 26: ...USER MANUAL 26...

Page 27: ...27 USER MANUAL...

Page 28: ...I A C E R Srl via Enzo Ferrari 2 30037 Scorz VE Italy www wepere com...

Reviews: