I-Tech WEPERE RESTART User Manual Download Page 10

USER MANUAL    

10 

Pack contents

The RESTART pack contains:
• 1 RESTART device;
• 1 medical power supply;
• 1 battery charger;
• 1 gel for ultrasound;
• 1 carry bag;
• 1 user manual.

Introduction to the technology

Ultrasound treatment represents a method that is based on the transfer of energy into the tissues 
which results in thermal and non-thermal biological effects.
This treatment is based on the contact between the head of the device and the tissues being 
treated. The optimal coupling of these two surfaces is ensured by the use of a water-based ultra-
sound gel. Through the contact described the transmission of general mechanical waves from 
the piezoelectric present inside the head of the device takes place. In order for there to be maxi-
mum transmission of energy from one medium to another, the impedance of the two must be the 
same. Obviously in the case of the human body, such condition hardly occurs.
The greater the impedance difference between the two media, the greater the reflection and con-
sequently the smaller the amount of energy that will be transferred. The impedance difference is 
greatest with aluminium-air interface, which is the first that ultrasound waves would have to over-
come to reach the body. To reduce this difference, a coupling substance must be used. If there is a 
small amount of air between the transducer and the skin, the proportion of ultrasound waves that 
would be reflected would reach 99.998%, i.e. there would be no transmission.
In addition to the phenomenon of reflection, if the wave does not hit the separation surface 
between the media at 90°, refraction occurs. In practice, the direction of the ultrasound beam 
through the second medium will be angled. The critical angle for ultrasound waves at the skin 
surface appears to be 15°.If the device head is oriented at a 15° angle to the surface, most of the 
beam will propagate through the epidermal tissues parallel to the skin rather than perpendicular 
to the tissues.
The absorption of the energy released by the ultrasound waves follows an exponential trend, in 
fact much more energy is absorbed in superficial tissues rather than in deep ones.
Precisely due to the trend of absorption, theoretically there is no point where all the energy is absor-
bed, but there is certainly a point where these levels are not sufficient to produce a therapeutic effect.
Generally tissues with the highest protein content will absorb more, unlike tissues with high water 
and low protein content, which will absorb a minimal amount of energy (blood and fat, for example).

Summary of Contents for WEPERE RESTART

Page 1: ...USER MANUAL Recover faster anytime anywhere...

Page 2: ......

Page 3: ...e description and controls 7 Labelling 8 Pack contents 10 How To Use Introduction to the technology 10 Contraindications 11 Warnings 12 Device use 12 Treatments 14 Looking after the device Maintenance...

Page 4: ...ns of the electromagnetic compatibility Directive 2014 30 UE of the European Parliament and of the Council of the 26th February 2014 to the current TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY...

Page 5: ...able anaesthetic mixture with air or with oxygen or with nitrous oxide device intended for continuous operation device not suitable for external use Intended purpose and scope of use The RESTART devic...

Page 6: ...0Hz 10 Duty cycle 5 50 100 mains powered 5 50 100 battery powered Power Can be set in 3 steps L low M medium H high Maximum power density 1 6W cm mains powered 0 8W cm battery powered Maximum output p...

Page 7: ...L Device description and controls 1 Treatment time indicator 2 Time button 3 Power adjustment button 4 Treatment power indicator 5 Battery indicator 6 Device ON OFF button 7 Emitting head 8 Power supp...

Page 8: ...Out 15V 1 2A Battery Ni MH AAA850mAh 4 8 V Ultrasound Carrier frequency of use 1MHz 10 Duty cicle 5 50 100 mains powered 5 50 100 battery powered Maximum output power 6 4W mains powered 3 2W batter po...

Page 9: ...documents accompanying the product IP22 Device protected against the penetration of solids with diameter d 12 5mm and against the fall of vertical drops of water when the device is held at 15 from the...

Page 10: ...sound waves would have to over come to reach the body To reduce this difference a coupling substance must be used If there is a small amount of air between the transducer and the skin the proportion o...

Page 11: ...advised not to wear metal objects to use ONLY the accessories supplied by the manufacturer It is forbidden to use the device in the presence of equipment for monitoring the user s vital functions equi...

Page 12: ...s used in strict compliance with the instructions given in this manual Should any foreign substances get into the device contact the dealer or manufacturer immedia tely Should the device fall check th...

Page 13: ...has been set and the head has been placed in contact with the skin the treatment will begin it is important to perform the treatment by continuously and uniformly moving the head around the ap plicat...

Page 14: ...manufacturer Treatment Intensity Minutes Muscle H Power supply Battery 15 Tendon Ligament L M Power supply M Battery 10 15 Aesthetic L Power supply L M Battery 5 10 REMEMBER TO keep the ultrasound hea...

Page 15: ...elative humidity from 15 to 93 pressure from 700 to 1060 hPa In the bag provided ambient temperature from 5 to 40 C relative humidity from 15 to 93 pressure from 700 to 1060 hPa Troubleshooting Any ty...

Page 16: ...ry indicator will flash from L M H in sequence Once char ging is complete the battery charge indicator will light up as H High WARNING the battery life cycle depends on the charge discharge cycles to...

Page 17: ...ided by IACER Shouldyouneedtoreturnthegoodsthenpleasepackthedeviceandalltheaccessoriessothatitwon t be damaged during transportation In order to be entitled to the warranty assistance the purchaser mu...

Page 18: ...facturer if non original spare parts are used the operation and safety of the product might be affected and the warranty will be null and void Interference and electromagnetic compatibility tables The...

Page 19: ...ns and to use caution in therapeutic applications on users with pacemakers However it is advisable to use the device keeping a distance of at least 3 metres from televisions monitors mobile phones or...

Page 20: ...USER MANUAL 20 RESTART All rights reserved RESTART and the logo are the exclusive property of I A C E R S r l and are registered trademarks Issue MNPG417 01 of 31 01 2022...

Page 21: ...21 USER MANUAL...

Page 22: ...omagnetic environment guidance RF emissions CISPR 11 Group 1 The RESTART uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interfere...

Page 23: ...line 2kV Line s to line 1kV Line s to line 2kV Line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variation...

Page 24: ...nsmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compl...

Page 25: ...ation distance according to the frequency of the transmitter m d 1 2 P from 150kHz to 80MHz d 1 2 P from 80MHz to 800MHz d 2 3 P from 800MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 01 0 38 0 38 0 73 1 1 2 1...

Page 26: ...USER MANUAL 26...

Page 27: ...27 USER MANUAL...

Page 28: ...I A C E R Srl via Enzo Ferrari 2 30037 Scorz VE Italy www wepere com...