IACER Srl
35
MNPG118-02
4.
4. Immediately after switching on, the device carries out a self-test
(around 10 seconds). At the end of the self-test display shows the last
performed treatment.
5.
Press
WAVEFORM MODE
(B1) button to show
: this icon indicates
ELECTROTHERAPY
treatment.
6.
Select waveforms by using the
knob
(3): TENS, EMS, IF-4P (Interferential 4
poles), IF-2P (interferential 2 poles), RUSSIAN (Kotz). The icon related to
the selected waveform starts flashing.
7.
Select the program (from P01 to P10) by pressing the
PROGRAM CC/CV
(B2) button: the symbol P (or S) of
the program starts flashing on the display. There are
two type of programs you can select: programs with
one phase (P) or with three phases (S). You can select
the program by keeping pressed
PROGRAM CC/CV
button for at least 5 seconds.
In the S programs display shows the phases total
number and the number of the phase you want to
set the parameters: press
B3-B7 buttons
and the
knob
(3) to select programs parameters
(contraction, recovery, width impulse, frequency,
etc.) for each phase (please see the following
instructions regarding each parameter). Press the
knob
to confirm the
value for each phase
8.
Select CONSTANT CURRENT (CC) or CONSTANT VOLTAGE (CV) by pressing
again
PROGRAM CC/CV
button. The unity of measurement changes from
mA (milliAmpere) to V (Volts).
9.
Select the waveform features by pressing
B3-B7 buttons
(each waveform
has different technical features):
•
B3: select frequency modulation/vector/Burst
•
B4: select frequency/carrier frequency
•
B5: select duty cycle/high frequency/width modulation
•
B6: select low frequency/impulse duration
•
B7: select therapy time/contraction-recovery cycle / slope
Press the button of the selected parameter and regulate the value by
using the
knob
(3).