24
www.i-sens.com
www.i-sens.com
25
Comparing the Control Solution Test Results
The test result of each control solution should be within the range
printed on the label of the test strip box. Repeat the control
solution test if the test result falls outside of this range. Out of
range results may occur in the following situations:
If results continue to fall outside the range printed on the test
strip box, the KetoSens Test Strip and Meter may not be working
properly. Do not use your system and contact your authorised
i-SENS sales representative.
Situations
• When the control solution bottle was not
shaken well,
• When the meter, test strip, or the control
solution were exposed to high or low
temperatures,
• When the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
• When the meter is not functioning
properly.
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
• When the control solution is
contaminated.
Do This
Repeat the
control solution
test by referring
to the
Note
on page 20.
Discard the
used
control solution
and repeat the
test using a new
bottle of control
solution.