Comparing the Control Solution Test Results
Repeat the control test if the test result falls outside the range
printed on label of the test strip vial. Out of range results may
occur due to the following factors:
• When the control solution vial was not shaken well,
• When the control solution is past its expiration date or is
contaminated,
• When the meter, the strip or the control solution were
exposed to high or low temperatures,
• When the first few drops of the control solution were not
discarded or the tip of the vial was not wiped clean,
• When the test strip is past its expiration date,
• When the meter is not functioning properly.
Note:
The CareSens Control Solution can be purchased separately.
Please contact your authorized i-SENS sales representative.
Caution:
To avoid infection when drawing a sample, use a lancet only
one time, and:
• Do not use a lancet that has been used by others.
• Always use a new sterile lancet.
• Keep the lancing device clean.
Note:
Repeated puncturing at the same sample site may cause pain
or skin calluses. Choose a different site each time you test.
Using the Lancing Device
You will need a lancing device in order to collect a blood sample.
You may use the lancing device contained in the CareSens N
Blood Glucose Monitoring System or any other medically
approved lancing device.
• The lancing device may not be used by more than one
individual. Ensure the lancing device is not shared among
different users.
• Use a soft cloth or tissue to wipe the lancing device.
If necessary, a small amount of alcohol on a soft cloth or
tissue may be used.
Adjustable Tip
Lancet Holder
Sliding Barrel
Release Button
Ejection System
Lancet