background image

18

   www.i-sens.com

www.i-sens.com  

19

Caution: 

If all the results you get on testing three times continuously are 

within the range printed on the test strip, the meter and test  

strips are working properly and you may use them for your blood 

glucose test.

Control Solution Testing

Step 2 

Shake the CareSens Control 

Solution vial before each test.

Remove the cap and squeeze the 

vial to discard the first drop. Then 

wipe the tip with a clean tissue 

or cloth. After the 

 symbol 

appears on the display, apply the 

solution to the narrow edge of the test strip until the meter 

beeps. Make sure the confirmation window fills completely.

Step 1 

Insert a test strip into the meter’s test strip port, 

with the contact bars facing upwards.

Gently push the test strip into the port until the 

meter beeps. Be careful not to break the strip 

while pushing it in.

The 

 icon will be displayed on the screen.

Step 3 

A test result will appear after the meter counts 

down from 5 to 1.

After your control solution result appears on 

the display, press   for 3 seconds till the ‘check’ 

icon appears on the display.

When the ‘check’ icon is displayed, the result 

is not stored in the meter’s memory and is not 

included in the 14-day averages.

Step 4

 

Compare the result displayed on the meter to 

the range printed on the test strip vial.

The result should fall within that range.

Used strips should be discarded safely in 

disposable containers.

Note:

 

The meter may switch off, if the blood sample is not applied 

within 2 minutes of the 

 icon appearing on the screen. 

If the meter turns off, remove the strip, reinsert, and start 

from step 1.

Caution: 

•    If the results you get are not within this range, the meter and 

test strip may not work properly. Then, stop using the meter 

and contact the nearest i-SENS representative.

•    The range printed on the test strip vial is for the CareSens 

Control Solution only. It does not have any connection to 

your blood glucose level.

Summary of Contents for CareSens N Voice

Page 1: ...c All Rights Reserved PGA1E3121 REVx 02 2012 The Biosensor Exppert i SENS Inc 465 6 Wolgye4 dong Nowon gu Seoul 139 845 Korea www i sens com Medical Technology Promedt Consulting GmbH Altenhofstrasse...

Page 2: ...stem The system provides you with safe convenient and painless blood glucose in vitro i e outside the body diagnostic monitoring You can obtain accurate results in just 5 seconds with a small 0 5 bloo...

Page 3: ...Lancing Device 21 Preparing the Lancing Device for Blood Sample Retrieval 22 Preparing the Meter and Test Strip 24 Flagging Post meal Test Results 24 Applying Blood Sample 25 Discarding Used Lancets 2...

Page 4: ...reSens N strip The average hematocrit range is between 30 and 50 However people with severe anemia and newborns may have higher hematocrit and this can affect the test results Very high 60 or more or...

Page 5: ...C 50 104 F Relative humidity 10 90 Hematocrit 20 60 CareSensNBloodGlucoseMonitoringSystem Components 01 CareSens N Blood Glucose Meter 02 Lancing Device 03 Lancets 10 04 Owner s Booklet 05 Quick Refe...

Page 6: ...ps past the expiration date Test strips in new unopened containers and test strips in containers that have been opened can be used up until the expiration date printed on the test strip box and contai...

Page 7: ...lts stored in the memory are displayed alarm appears when the time alarm has been set mmol L unit for measuring blood glucose Battery icon indicates meter battery is running low and needs to be replac...

Page 8: ...adjust until the correct year appears Press and hold button to scroll through the numbers quickly After setting the year press the S button to confirm your selection and enter the month setting mode S...

Page 9: ...is case the unit can not be changed and the unit change setting step will be excluded Step 9 On pressing or the screen will display the On or OFF Press the S button to confirm the selection The meter...

Page 10: ...e and is used to check that the meter and the test strips are working properly The test strip vials have CareSens Control Solution ranges printed on their labels Compare the result displayed on the me...

Page 11: ...displayed on the screen Step 3 A test result will appear after the meter counts down from 5 to 1 After your control solution result appears on the display press for 3 seconds till the check icon appe...

Page 12: ...e contact your authorized i SENS sales representative Caution To avoid infection when drawing a sample use a lancet only one time and Do not use a lancet that has been used by others Always use a new...

Page 13: ...Sample Retrieval Step 4 Select a desired depth of one to five 1 5 on lance s adjustable tip Rotate ring to align desired number with arrow A beginning setting of three 3 is recommended Step 5 To cock...

Page 14: ...meter s test strip port Push the strip in gently until the meter beeps Be careful not to bend the test strip The symbol will appear on the screen Step 7 If you want to attach an post meal flag to a t...

Page 15: ...ically switch off after 3 seconds Discard used test strips safely in disposable containers Step 1 Unscrew general lancing device tip Step 2 Place protective cover on lancet Push the lancet ejector for...

Page 16: ...observed during testing AlternativeSitesforTesting Things to know when using AST Please understand the following things before testing outside of the fingertip arms palms thighs and calves The capilla...

Page 17: ...splayed even though you do not have hyperglycemia or hypoglycemia AST Precautions Do not ignore the symptoms of hyperglycemia or hypoglycemia When the results of the test do not reflect one s opinion...

Page 18: ...the last 14 days ViewingTest Results Stored in the Meter s Memory Step 2 Press to view the average value and the number of tests performed before eating a meal for the last 14 days On pressing again t...

Page 19: ...will be displayed The screen will then automatically change to the memory check mode At this time the bell icon indicating that the PP2 alarm has been set will be displayed on the screen Step 2 Setti...

Page 20: ...ncel the alarm1 The OFF icon will blink on the screen Step 3 Press to adjust the time of the alarm1 A number representing the time will blink on the screen Press to set the time Press to end Setting t...

Page 21: ...nto place Note Removing the meter batteries will not affect your stored result However you may need to reset your meter settings See pages 12 14 Caring for Your System Use a soft cloth or tissue to wi...

Page 22: ...40 C 50 104 F and repeat the test after 30 minutes Understanding Error and Other Messages Note If the error messages persist contact your authorized i SENS sales representative Message What It Means...

Page 23: ...r authorized i SENS sales representative Performance Characteristics The performance of CareSens N Blood Glucose Monitoring System Strips has been evaluated in laboratory and in clinical tests Accurac...

Page 24: ...owsthattherecouldbevariationofupto4 2 Warranty Information Manufacturer sWarranty i SENS Inc warrants that the CareSens N Meter shall be free of defects in material and workmanship in normal use for a...

Reviews: