17
-
1.
Take a test strip from the test strip vial
and close the cap quickly and tightly.
2.
Insert the test strip into the test strip slot,
and open 1 level of Control Solution.
3.
Squeeze a drop of control solution onto a
plain nonabsorbent plastic/surface, and
close the control solution vial immediately.
4.
Apply the control solution drop to the
sample collection area of the test strip.
The solution is automatically drawn in
and a “beep” sound is heard for
confirming.
5.
The measurement appears after 6
seconds. Compare it with the range
printed on the Strip vial.
NOTE:
DO NOT APPLY THE CONTROL SOLUTION DIRECTLY TO
THE TEST STRIP! To avoid contaminating the control
solution with the substance of the test strip.
In the case that the control measurements fall outside of
the acceptable range, please call customer service for
assistance.