6. CLEANING AND MAINTENANCE
7. OPERATING AND STORAGE CONDITIONS
6.1. CLEANING
6.2. MAINTENANCE
6.1.1. I-PEN
®
DEVICE
6.1.2. SINGLE-USE-SENSORS
6.2.1. TROUBLESHOOTING
The I-PEN
®
can be cleaned with a damp cloth or alcohol wipe as required. When cleaning, it is
important to keep the electronic contacts of the control unit and Reader dry. The electronic contacts
and docking port should also be kept free of dust and dirt.
To ensure warranty coverage and reliable system operation, defective system components should
be serviced and/or replaced exclusively by I- MED Pharma Inc. authorized personnel, and replacement
parts should be those specified by the manufacturer. It is important to use and store the device within
the environmental conditions shown in the table below.
Single-Use-Sensors are for use on a single eye. Never reuse or try to clean a Single-Use-Sensor.
Single-Use-Sensors are for single use only. Never reuse or try to clean a Single-Use-Sensor.
The I-PEN
®
Osmolarity System is designed to work without direct service or preventive maintenance.
If quality checks fail, contact I-PEN
®
Customer Support.
Battery should be replaced when “Low Bat” indication appears on the screen.
Condition
Possible Cause
Recommended Action
The “I-PEN Ready” message
does not display
Battery not installed.
Device malfunction.
Verify that the correct type of battery is
installed, and that the battery is fresh.
Contact I-PEN
®
customer support.
A “beep” is not heard when
the device is turned on
Device malfunction.
Contact I-PEN
®
customer support.
“Low Bat” indication appears
on screen
Battery close to
end-of-life.
Replace battery.
Screen goes dim
Battery close to
end-of-life.
Replace battery.
Transport and Storage Temperature
2°–35°C/36°–95°F
Transport and Storage Relative Humidity
10–85% non- condensing
Transport and Storage Altitude
0–2,000 meters
Operating Temperature
15°– 30°C/59–86°F
Operating Altitude
0–2,000 meters
Operating Relative Humidity
10–85% non- condensing
NOTE:
If the Recommended Action does not solve the problem,
contact I- MED Pharma Inc. Customer Support.
Single-Use-Sensors may be ordered on line at www.imedpharma.com or by calling your
representative at I- MED Pharma Inc. Tel. Number: (800) 463- 1008 or (514) 685- 8118.
I-PEN
®
User Guide
|
CLEANING AND MAINTENANCE
I-PEN
®
User Guide
|
OPERATING AND STORAGE CONDITIONS | TECHNICAL SPECIFICATIONS
8. TECHNICAL SPECIFICATIONS
Calibration
No calibration required
Degree of Protection Against Electric Shock
BF Type applied part
Size (not including probe holder)
W 140mm
L 223mm
H 140mm
Weight
50 gm
Battery
CR2032
Frequency
80 Hz
Peak Voltage
± 1.5V
Current Source
Max100 μA AC
Sinus Distortions
5%
WARNING:
Cleaning fluids should never be used on the Single-Use-Sensor.