Push the Ejector button and the Single-Use-Sensor 

will be ejected.

You may now discard the 

Single-Use-Sensor.

The LCD will display a test result.

See the next Section for a discussion of how to 

interpret the measurement results.

4.6. EJECT THE SINGLE-USE-SENSOR

5.  EXPECTED RESULTS

I-PEN

®

 test results are displayed on the LCD in units of mOsms/L. No calculations are required. The 

chart below shows some typical test results and their possible interpretation. All such interpretations are 

subject to the review of the physician or other medical professional.

I-PEN

®

 User Guide  

|

  PERFORMING AN OSMOLARITY MEASUREMENT

I-PEN

®

 User Guide  

|

  EXPECTED RESULTS

1. 

Do not immerse the tip of the Single-Use-Sensor in the lower tear meniscus.

2. If the reading on the LCD screen shows an “Error”, you may attempt another reading in the same eye by   

  pressing the Ready button.

4.5.1. TIPS FOR USE

4. Approach at a 30-45 degree angle from horizontal and gently lower the end of the Single-Use- 

 

  Sensor on to the conjunctiva on the inside of the lower eyelid. 

5. When correctly placed, the tip of the Single-Use-Sensor should be depressing the surface slightly    

  so that both gold nodes are in good contact with the conjunctiva.

6. The I-PEN

®

 will take make an audible beep after several seconds and display the reading

  on the LCD screen.

4.5. TAKING A READING

1. Ask the patient to gently squeeze their eyelids shut  

  for 30-60 seconds prior to taking a reading.

2. Position the tip of the disposable Single-Use-Sensor  

  just above the lower eyelid with the LCD screen  

  facing upwards.

3. Turn on the I-PEN

®

 as indicated in section 4.4 

only  

  when you are ready to take the reading

 by pushing  

  the on/off switch to the on position.

4.4. TURN ON THE DEVICE

Push the On/Off switch 

6

. You should hear a 

“beep” and the LCD display 

7

 should display 

the “I-PEN Ready” message.

CAUTION: 

Replace the device if a “beep” is not heard after turning it on.

NOTE:

 In order to conserve battery life, the I-PEN

®

 is programmed to enter Sleep Mode 

automatically thirty seconds after it powers up. In doing so, this can invalidate the SUS and 

a new SUS should be inserted.

CAUTION: 

Do not use the unit if the Reader does not display “I-PEN Ready”.

DRY EYE SEVERITY SCALE

380

365

350

335

320

305

290

275

Normal

< 290

Marginal

290-310

Mild

310-330

Moderate

330-350

Severe

> 350

5.1. REFERENCE TEAR OSMOLARITY VALUES

Reading in mOsms/L 

(Use result from eye with 

highest reading)

Variance Between Right & 
Left Eye

Interpretation

<290

Normal Patient

290-310

≤7

Normal Patient

290-310

≥8

Dry Eye Disease Patient

>310

Dry Eye Disease Patient

Summary of Contents for I-PEN

Page 1: ...FOR QUANTITATIVE MEASUREMENT OF OSMOLARITY OF OCULAR TISSUES Osmolarity System USER MANUAL osdcare com...

Page 2: ...NDITIONS 13 8 TECHNICAL SPECIFICATIONS 13 9 ELECTROMAGNETIC EMISSIONS 14 10 ELECTROMAGNETIC IMMUNITY 15 10 1 Recommended Separation Distances 17 10 1 1 Applicable Standards 17 11 LABELS AND SYMBOLS 18...

Page 3: ...smolarity Single Use Sensor in conjunction with the I PEN Osmolarity System provides a quick and simple method for determining tear osmolarity using impedance measurements of the saline concentration...

Page 4: ...WARNING Do not test patients who have used eye drops within two hoursprior to testing WARNING Do not test patients wearing makeup on eyelids WARNING Do not test patients within 10 minutes after remova...

Page 5: ...n 5 insert the disposable Single Use Sensor CAUTION Replace the device if a beep is not heard after turning it on CAUTION It is important to visually inspect the Single Use Sensor before use In the ca...

Page 6: ...ontact with the conjunctiva 6 The I PEN will take make an audible beep after several seconds and display the reading on the LCD screen 4 5 TAKING A READING 1 Ask the patient to gently squeeze their ey...

Page 7: ...sage does not display Battery not installed Device malfunction Verify that the correct type of battery is installed and that the battery is fresh Contact I PEN customer support A beep is not heard whe...

Page 8: ...s RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic em...

Page 9: ...be necessary such as re orienting or relocating the I PEN 10 1 RECOMMENDED SEPARATION DISTANCES 10 1 1 APPLICABLE STANDARDS Recommended separation distances between portable and mobile RF communicati...

Page 10: ...requirements and standards and are briefly reviewed below REF 600 SN nn xxxx mm yyyy LifeCare Ltd 2 Zipori Street Tiberias 1424602 Israel PRT 14 200 Rev 01 CAUTION At the end of its useful life the sy...

Page 11: ...I MED Pharma Inc 1601 St Regis Blvd Dollard des Ormeaux QC Canada H9B 3H7 Tel 514 685 8118 Toll free 800 463 1008 Fax 514 685 8998 info imedpharma com SEE THE DIFFERENCE Product IPUMEN 0718...

Reviews: