OF OSMOLARITY OF OCULAR TISSUES

Osmolarity System

USER MANUAL

osdcare.com

Summary of Contents for I-PEN

Page 1: ...FOR QUANTITATIVE MEASUREMENT OF OSMOLARITY OF OCULAR TISSUES Osmolarity System USER MANUAL osdcare com...

Page 2: ...NDITIONS 13 8 TECHNICAL SPECIFICATIONS 13 9 ELECTROMAGNETIC EMISSIONS 14 10 ELECTROMAGNETIC IMMUNITY 15 10 1 Recommended Separation Distances 17 10 1 1 Applicable Standards 17 11 LABELS AND SYMBOLS 18...

Page 3: ...smolarity Single Use Sensor in conjunction with the I PEN Osmolarity System provides a quick and simple method for determining tear osmolarity using impedance measurements of the saline concentration...

Page 4: ...WARNING Do not test patients who have used eye drops within two hoursprior to testing WARNING Do not test patients wearing makeup on eyelids WARNING Do not test patients within 10 minutes after remova...

Page 5: ...n 5 insert the disposable Single Use Sensor CAUTION Replace the device if a beep is not heard after turning it on CAUTION It is important to visually inspect the Single Use Sensor before use In the ca...

Page 6: ...ontact with the conjunctiva 6 The I PEN will take make an audible beep after several seconds and display the reading on the LCD screen 4 5 TAKING A READING 1 Ask the patient to gently squeeze their ey...

Page 7: ...sage does not display Battery not installed Device malfunction Verify that the correct type of battery is installed and that the battery is fresh Contact I PEN customer support A beep is not heard whe...

Page 8: ...s RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic em...

Page 9: ...be necessary such as re orienting or relocating the I PEN 10 1 RECOMMENDED SEPARATION DISTANCES 10 1 1 APPLICABLE STANDARDS Recommended separation distances between portable and mobile RF communicati...

Page 10: ...requirements and standards and are briefly reviewed below REF 600 SN nn xxxx mm yyyy LifeCare Ltd 2 Zipori Street Tiberias 1424602 Israel PRT 14 200 Rev 01 CAUTION At the end of its useful life the sy...

Page 11: ...I MED Pharma Inc 1601 St Regis Blvd Dollard des Ormeaux QC Canada H9B 3H7 Tel 514 685 8118 Toll free 800 463 1008 Fax 514 685 8998 info imedpharma com SEE THE DIFFERENCE Product IPUMEN 0718...