1 Overview
2
!
Start/Stop-button, described in text
as left button
@
Menu-button, described in text
as right botton
#
Infrared-eye to transmit the
stored measured data to the
doctor’s computer
$
Cuff
%
Battery compartment on the back
^
Display
!
!
@
@
#
#
$
$
%
%
^
^
11
6 Descriptions Of Errors And How To Correct Them
6.2 Trouble Shooting
Description of error
The measurement cannot be stopped and the
cuff does not deflate instantly when pushing
the left button once
Left button has been pushed but the measure-
ment does not start (pumping stops early)
Display shows no symbols
Display shows battery symbol
Time and/or date wrong
Frequent error messages
Correction of error
Open cuff, remove batteries, stop using
the device. Contact Service!
Check batteries, change if necessary
Check batteries, change if necessary
Change both batteries
see Setting Time And Date
a) put on cuff correctly (see 4.2)
b) wrist too small or big 14 cm (5.5“)
to 19.5 cm (7.7“) circumference
c) hold arm and fingers still during
measurement