HyQuest Solutions Pty Ltd
WS200/WS300/WS500
Page
10 of 28
©
Copyright
-
Damaged brush or terminal wire
-
Wiring to slipring break
-
Bridge slipring to winch
frame. If signal emits
that means proper
circuit
between brush
and red
terminal. If no
signal,
loose red
terminal
connection,
broken, or
brush
sticking. To check
condition of brush
remove signal adaptor
by unscrewing the black
terminal and the
mounting screw. Ensure
that brush and spring
move freely in signal
adaptor housing or
repair with file and
extend the spring.
-
Check the wire on the
slipring if loose or
broken. If broken see
section fitting
ame
r
graph
cable to
slipring.
Fault Symptom
Possible Cause
Action
Current meter counter
registering continuously
-
Faulty lead
-
Fit lead to counter only.
Signal shows internal
short circuit in lead.
Inspect for break or
damage.
-
Current meter or angle plug are
faulty
-
Disconnect meter. If
signal still emitting
continuously. Check the
earth inside angle plug
or damage to cable
between meter and
gauging weight. If
signal stops emitting
continuously repair the
current meter.
-
Meter suspension cable is
damaged, or slipring wires
earthing to drum end
-
With lead connected to
winch, bridge red
terminal to frame. If no
signal terminal loose or
corroded.
ISSUE 3:11/6/19