Hyquest Solutions Pty Ltd

 

Top Set Rod – Model: MagnaRod 

© Copyright 

 

 

 

 

- 13 -

   

 

Issue 1.03   5 Mar, 2013 

manner.    All  serial  data  sent  from  the 

MagnaRod

  will  be  transmitted  via 

Bluetooth

®

.    The 

connection  can  be  disengaged  by  selection  in  the  ‘

Bluetooth

’  menu.  The  operator  must 

monitor the integrity of the connection from the external device.  If the connection becomes 
broken, the 

MagnaRod

 will revert back to 

Bluetooth

®

 

DISCOVER mode.  The blue flashing 

LED  will  alert  the  operator  of  a  disconnection  between  the 

MagnaRod

  and  the  external 

device.    Connection  will  have  to  be  rediscovered  by  the  external  device.    Any  active 
measurement  occurring  during  a  broken  communications  link  will  terminate  and  the 
measurement  will  need  to  be  retaken.  The 

MagnaRod

  will  not  automatically  power  down 

when 

Bluetooth

®

 is enabled and connected.   

 

4.7.1  Serial Communications Interface 

 

The 

MagnaRod

  can  be  connected  to  a  hand-held  or  notebook  computer  or  any  device  that 

can transmit  and  receive serial  communications  using 

Bluetooth

®

  using  an application such 

as  HyperTerm.  Commands  can  be  sent  to  the 

MagnaRod 

to  perform  special  functions  or 

select  operating  modes.      The  following  describes  the  serial  data  requirements,  commands, 
and data formats that are used for the serial communications interface : 

 

 

Baud Rate :   

19,200  

 

Format : 

 

8 data bits, no parity, 1 stop bit 

 

Flow Control : 

None 

 

The  following  list  shows  the  commands  and  functions  that  can  be  performed  through  the 
serial communications interface. 

 

Command

 

 

Function Performed

 

 

S

 

 

Start a measurement 

 

T

 

 

Terminate the measurement 

 

I

 

 

Abort the measurement 

 

R

 

 

Resend the last data string (** not implemented yet **) 

 

U

 

 

Buzzer ON 

 

Z

 

 

Buzzer OFF 

 

H

 

 

Cat Whisker Speed Select, set to Fast 

 

L

 

 

Cat Whisker Speed Select, set to Slow 

 

M

 

 

Select Magnetic Head meter processing 

 

C

 

 

Select Cat Whisker Head meter processing (Fast) 

 

 

Start a measurement without performing a calibration 

 

Q

 

 

Start a measurement with no time limit (continuous measure) 

 

V

 

 

Software Version

 

NOTE:  Commands sent to the MagnaRod are case sensitive – and must be upper case only. 

  When  the  commands  ‘

S

’,  ‘

T

’,  and  ‘

I

’  are  received  by  the 

MagnaRod

  from  an 

externally  connected  device,  the 

MagnaRod 

sends  an  acknowledge  character  ‘

A

’ 

back  to  the  device.    For  the  ‘

S

’  command,  the 

MagnaRod

  sends  acknowledge 

Summary of Contents for MagnaRod MAGD/CMC

Page 1: ...C HYQUEST SOLUTIONS PTY LTD PO BOX 332 LIVERPOOL B C NSW 1871 AUSTRALIA Phone Int 612 9601 2022 Fax Int 612 9602 6971 Phone Nat 02 9601 2022 Fax Nat 02 9602 6971 Email sales hyquestsolutions com au We...

Page 2: ...4 Current Meter Counter Operation 8 4 5 Current Meter Counter Menus 9 4 6 Indicator LEDs 12 4 7 Bluetooth Communications 12 4 7 1 Serial Communications Interface 13 4 7 2 Data Conversion Example 15 4...

Page 3: ...quations and 3 rotations sec ranges The MagnaRod counter operates with 3 x AA alkaline 1 5V batteries and comes in a robust waterproof enclosure The commands are identical to the CMCsp commands so exi...

Page 4: ...nal to determine if a contact closure has occurred and if successive signals are of the same characteristics The MagnaRod then determines contact closure conditions and either counts the signal closur...

Page 5: ...ress and hold both buttons to turn on the MagnaRod Power OFF Press and hold both buttons again to turn off the MagnaRod Left Right Button Button Bed TSR Surface Stop CMC Start Up Dn Menu Select Hold T...

Page 6: ...being measured from the surface down to the current meter The total water depth or Sounding is displayed on the bottom of the screen Depth too small indicates you have moved the rod less than 0 02ft 5...

Page 7: ...e from the river bed to the current meter shaft when the Top Set Rod is in the lowest position 3 LCD Options allows 3 more LCD selections to be chosen The contrast can vary between 6 light and 24 dark...

Page 8: ...s pulses are detected 3 When the elapsed timer exceeds the measurement interval and a meter pulse is detected the timer will be stopped The final count elapsed time and the calculated velocity are all...

Page 9: ...contacts this becomes extremely important since Cat s Whisker contacts offer the poorest quality for marking and counting rotations In Cat s Whisker mode there are 2 speed settings FAST is selected fo...

Page 10: ...tions on the screen match the equations on your Current Meter calibrations sheets 6 Select Save and Return to go back to the previous menu 7 When More Settings Test is selected 5 more selections becom...

Page 11: ...e for the magnetic head contact does not apply but can be performed 11 The Spin Test is an elapsed time measurement from the time a meter is spun to when rotation of the meter stops Press Start and th...

Page 12: ...d IEEE 802 14 that offers a transparent serial port protocol for communication When the MagnaRod is turned On the MagnaRod determines if Bluetooth is installed If available Bluetooth can be enabled an...

Page 13: ...n such as HyperTerm Commands can be sent to the MagnaRod to perform special functions or select operating modes The following describes the serial data requirements commands and data formats that are...

Page 14: ...t that is transmitted to the externally connected device can be used to display elapsed time and velocity while the measurement is being performed At the end of the measurement interval or if the T co...

Page 15: ...uously output a 1 second data display without a time limit The T or I command must be used to terminate the operation V The software version of the MagnaRod is sent in the following manner vX X Exampl...

Page 16: ...terval time is then used to compute the measured velocity For the Normal mode the 1 second output data format displays the accumulating time interval counts based on a 300 Hz sample rate The following...

Page 17: ...interface Command Function Performed N Spin Test S Start Spin Measurement subcommand in Spin Test A Stop Spin Measurement subcommand in Spin Test U Buzzer ON Z Buzzer OFF M Magnetic Head Meter C Cat...

Page 18: ...ove the batteries and any remaining water Allow the unit to dry out completely before reinstalling the batteries There are no user serviceable parts inside the MagnaRod If the unit does not function p...

Page 19: ...porary immersion Note 60 1 Meter cup width for 6 cups Water Velocity Flow absolute maximum 1 Normal Mode Price AA meter Mag 45 ft s 13 7m s 99o to 204o dwell CatW 22 ft s 6 7m s 10o to 62o dwell Pygmy...

Page 20: ...er Head 1 NOTES a Measurements and Computations based on the following Rating Table Equations Price AA Meter V 2 2048 0 0178 where V Velocity ft s T T Seconds rotation Pygmy Meter V 0 9604 0 0312 T b...

Page 21: ...f the spin test the MagnaRod sends a final data string that displays the total number of counts and total time all shown in decimal notation This format makes it easy to copy the final total number of...

Page 22: ...nds to the total number displayed 3 The maximum number of measured meter contact closures clicks is 999 If the accumulating contact closures continues beyond the maximum data form the count rolls over...

Page 23: ...in the Accessories folder listed in the Programs list on the desktop Within HyperTerminal and before spinning the meter select Capture Text found in the Windows label Transfer tab When the Capture Tex...

Page 24: ...Do not import column skip in the Column data format window This will show Skip in the column 1 heading 11 Click on column 3 where the Time data is shown in HEX ASCII and change in the Column data form...

Page 25: ...response of a current meter created from data in a Spin Test file Spin Test for Meter 1234567 0 10 20 30 40 50 60 70 80 90 100 110 120 130 1 6 11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96 101...

Page 26: ...wing simple formula Dwell Angle Contact time x 360 Total Time The contact time and total time shown above in the formula are expressed in terms of the number of samples measured at a sample rate of 12...

Page 27: ...e 27 4 06 INSPECTED __________________ Anthony Gemayel Mechatronics Engineer _____________________________________________________________ The above meter has been calibrated in accordance with AS 377...