background image

4

LABELS 

The following labels and symbols appear on the control unit, garments, and/or packaging.

Symbol

Description

Location

IP21

Degree of protection against ingress of water

On base of control unit

Read instructions before use

On base of control unit 

and attachment tag

Level of protection type BF equipment

On base of control unit

Double insulation

On power adapter

Direct current

On base of control unit

Alternating current

In manual

Manufacturer’s name and address

On base of control unit 

and attachment tag

Date of manufacture

On base of control unit

Separate collection for waste electrical and electronic equipment 

On base of control unit

Serial number of the console

On base of control unit

Fragile, handle with care

On package

Keep dry 

On package

This side up 

On package

Keep away from sunlight 

On package

93%RH

10%RH

Transportation & storage humidity limitation 

On package

1060hPa

700hPa

Transportation & storage atmospheric pressure limitation 

On package

-4ºF

-20ºC

70ºC
158ºF

-4ºF

-20ºC

70ºC
158ºF

Transportation & storage temperature limitation 

On package

Place in and out of standby mode

On top of control unit

Summary of Contents for Normatec 2.0

Page 1: ...Normatec 2 0 Operating Instructions Instrucciones de operaci n...

Page 2: ...such as near a bathtub kitchen sink laundry tub or swimming pool To avoid damage and risk of electric shock never spill liquid of any kind on the system Do not place the system power supply or any acc...

Page 3: ...duct if you are experiencing inflammation an infection pain of unknown origin bleeding internal or external at or near the site of application or if you have a wound at or near the site of application...

Page 4: ...and address On base of control unit and attachment tag Date of manufacture On base of control unit Separate collection for waste electrical and electronic equipment On base of control unit Serial numb...

Page 5: ...sons be presented on the medical control unit itself In manual Tip to provide guidance to make use easier Risk to user is considered negligible In manual INDICATIONS FOR USE The Normatec 2 0 System is...

Page 6: ...Pressure adjustment button increase 6 Pressure adjustment button decrease 7 Display screen 8 Zone Boost button 9 Play pause button Normatec 2 0 Hose single person use only 1 Junction box air outlets 2...

Page 7: ...S CONTROL UNIT Read all warnings at the beginning of this manual If you do not understand these operating instructions contact Hyperice at 1 949 565 4994 SET UP THE SYSTEM Step 1 Plug the power supply...

Page 8: ...control unit is on the green LED next to the power button will be lit up ADJUST THE INTENSITY Adjust the intensity level of the session by pressing the level adjustment buttons on the left and right o...

Page 9: ...lling away TURN OFF THE CONTROL UNIT To turn off the system press the power button and confirm that the green power LED is off USE ZONE BOOST During your session you can increase the intensity of a si...

Page 10: ...during a massage each zone of the attachments will first compress in a pulsing manner and then release as the compression pattern works its way up your limb When the top zone completes its massage th...

Page 11: ...on regarding available replacement parts and accessories TECHNICAL INFORMATION Do not attempt to take apart the system The system has no user serviceable parts There are no user replaceable fuses BLUE...

Page 12: ...icables aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l utilisateur de l appareil doit accepter...

Page 13: ...d provide reasonable assurance of such This information may not be applicable to other medical electrical equipment older equipment may be particularly susceptible to interference GENERAL NOTES Medica...

Page 14: ...rposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker IEC 61000 3 3 Complies ELECTROMAGNETIC IMMUNITY This equipment is intended for use in the electromagnetic environment speci...

Page 15: ...requency magnetic field 50 60 Hz IEC 61000 4 8 30 A m 30 A m Equipment that emits high levels of power line magnetic fields in excess of 3A m should be kept at a distance to reduce the likelihood of i...

Page 16: ...tion distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may...

Page 17: ...applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the sepa...

Page 18: ...junction box Check that the attachments do not have an air leak The system stopped pumping The hose is not securely connected The attachments have been damaged Check that the hose is securely connect...

Page 19: ...he attachments and other accessories collectively the product are not used in accordance with the user manual are otherwise misused or modified in any way and or are repaired or altered by anyone othe...

Page 20: ...changes must have a Return Merchandise Authorization RMA number Obtain an RMA number by contacting us at 1 949 565 4994 or customersupport hyperice com Returned items without an RMA number will not be...

Page 21: ...del agua como la ba era el fregadero de la cocina el lavadero o la piscina Para evitar da os y riesgos de descargas el ctricas nunca derrame l quido de ning n tipo en el sistema No coloque el sistema...

Page 22: ...ece una inflamaci n infecci n dolor de origen desconocido o sangrado interno o externo o si presenta una herida en el punto de aplicaci n o cerca de este Consulte con su m dico antes de utilizar este...

Page 23: ...e control Corriente alterna En el manual Nombre y direcci n del fabricante En la base de la unidad de control y en la etiqueta de la funda Fecha de fabricaci n En la base de la unidad de control Recog...

Page 24: ...btener informaci n importante sobre las se ales de alerta como advertencias y precauciones que no pueden por diversas razones presentarse en la unidad de control m dico en s misma En el manual Consejo...

Page 25: ...de la presi n aumentar 6 Bot n de ajuste de la presi n disminuir 7 Pantalla 8 Bot n de intensificaci n por zonas 9 Bot n de Reproducir Pausar Tubo flexible Normatec 2 0 para el uso de una sola person...

Page 26: ...ONTROL Lea todas las advertencias incluidas en este manual En caso de no entender estas instrucciones de uso p ngase en contacto con Hyperice llamando al 1 949 565 4994 CONFIGURACI N DEL SISTEMA Paso...

Page 27: ...al bot n de encendido AJUSTE DE LA INTENSIDAD Para ajustar el nivel de intensidad de la sesi n pulse los botones de ajuste de presi n situados a la izquierda y a la derecha del indicador de nivel El...

Page 28: ...paga el LED de encendido verde USO DE ZONE BOOST INTENSIFICACI N POR ZONAS Durante la sesi n es posible incrementar la intensidad de una zona con la funci n Zone Boost Intensificaci n por zonas La int...

Page 29: ...amiento y los golpecitos que se realizan durante un masaje cada segmento de la funda ejerce primero una compresi n pulsada y luego la detiene a medida que el programa de compresi n va subiendo por la...

Page 30: ...te nuestro sitio web hyperice com para obtener informaci n sobre las piezas de repuesto disponibles INFORMACI N T CNICA No intente desmontar el sistema ya que no incluye piezas que pueda reparar el us...

Page 31: ...ado de la unidad de control Le pr sent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1...

Page 32: ...nto del dispositivo sino que lo aseguran en la medida de lo razonable Esta informaci n puede no resultar aplicable a otros equipos electrom dicos De hecho los equipos m s antiguos pueden resultar espe...

Page 33: ...ctuaciones de tensi n parpadeo IEC 61000 3 3 Conforme INMUNIDAD ELECTROMAGN TICA Este equipo est dise ado para su uso en el entorno electromagn tico descrito a continuaci n El usuario del equipo deber...

Page 34: ...0 Hz IEC 61000 4 8 30 A m 30 A m Aquellos equipos que emiten niveles altos de campos magn ticos de red con un exceso de 3 A m deber n mantenerse a cierta distancia para reducir la probabilidad de inte...

Page 35: ...os de los transmisores de radiofrecuencia fijos conforme a lo determinado mediante un estudio del lugara deber ser menor que el nivel de conformidad en cada intervalo de frecuenciab Pueden producirse...

Page 36: ...icable a la frecuencia del transmisor donde P es la potencia de salida m xima del transmisor en vatios W seg n el fabricante del transmisor Nota 1 A 80 MHz y 800 MHz se aplicar la distancia de separac...

Page 37: ...fuga de aire El sistema ha dejado de bombear El tubo flexible no se ha conectado correctamente Las fundas se han da ado Compruebe que el tubo flexible est correctamente conectado a la unidad de contr...

Page 38: ...o la mano de obra y b sustituir la unidad de control por una del mismo tipo o por el dise o m s reciente en dicha fecha Las garant as anteriores no cubren el uso y desgaste normal ni los da os est ti...

Page 39: ...a comprado fundas de talla incorrecta podr cambiarlas en un plazo de catorce 14 d as a partir de la fecha de compra Todos los cambios y devoluciones estar n sujetos a las condiciones enunciadas a cont...

Page 40: ...hyperice com Copyright 2020 Hyperice Inc All rights reserved NormaTec Industries LP 480 Pleasant Street Suite A200 Watertown Massachusetts 02472 USA 70 00046 Rev 1...

Reviews: