15
RF Proximity
(IEC 61000-4-3)
385 Pulse Modulation, 18 Hz
450 FM + 5Hz deviation: 1 kHz sine
710
745
780 Pulse Modulation: 217 Hz
810
870
930 Pulse Modulation: 18 Hz
1720
1845
1970 Pulse Modulation: 217 Hz
2450 Pulse Modulation: 217 Hz
5240
5500
5785 Pulse Modulation: 217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
Equipment with high RF emissions
should be kept at a distance to reduce
the likelihood of interference.
Surges on AC
mains lines
(IEC 61000-4-5)
Common Mode: ± 2 kV
Differential Mode: ± 1 kV
± 2 kV
± 1 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency
magnetic field
50/60 Hz
(IEC 61000-4-8)
30 A/m
30 A/m
Equipment that emits high levels of
power line magnetic fields (in excess of
3A/m) should be kept at a distance to
reduce the likelihood of interference.
Voltage dips
and short
interruptions on
AC mains input
lines
(IEC 61000-4-11)
Dip > 95%, 0.5 cycles
Dip > 95%, 1 cycle
Dip 30%, 25 cycles (50 Hz)
30 cycles (60 Hz)
Drops >95%, 250 cycles (50 Hz)
300 cycles (60 Hz)
0.5 cycles
1 cycle
25 cycles (50 Hz)
30 cycles (60 Hz)
250 cycles (50 Hz)
300 cycles (60 Hz)
Mains power should be that of a typical
commercial or hospital environment. If
you require continued operation during
power mains interruptions, ensure that
batteries are installed and charged.
Ensure that battery life exceeds
longest anticipated power outages
or provide additional uninterruptible
power source.