18
If you wish to perform a QC test after a certain amount of measurements,
a QC set consisting of a special QC cassette and a corresponding QC
configuration on the RFID card can be provided for this purpose (please
contact your supplier).
When a QC test is required, please take the provided QC set and insert
the QC cassette into the cassette adapter according to the
Test Insertion
point of the measurement procedure and place the reader on top. Briefly
press the button and
RFID
will be displayed. Put the QC RFID card on top
of the reader and wait until the QC configuration data has been uploaded
(audible signal). Alternatively, select the QC test from the reader memory.
Briefly press the button again to start the QC test. The QC test result will be
displayed afterwards.
Attention:
The result of the QC test may be
PASS
or
FAIL
.
The reader will be able to measure again after the QC test,
even if the QC result was
FAIL
. When a QC test fails, the
reader should not be used for measurements because
results could be faulty. The Cube should be switched
off and on again, and the bottom part of the Cube that
performs the reading should be cleaned. If the problem
persists, please contact your supplier for a replacement.