Hydro-View User Guide HD0531 Rev 2.0.0 61
Figure 50: Manual Averaging
Touch Start when the sample collection starts and touch Stop when collection ceases
(Figure 51).
Figure 51: Manual Averaging Started
When averaging completes, a message is displayed giving the option of adding a new
point to the Calibration (Figure 52).
Figure 52: Manual Averaging Stopped
If a valid sample has been taken for a Laboratory measurement, then touch Yes. A new
point is added to the list and highlighted in red. Once a Laboratory test result for the
relevant batch is complete, the true moisture value can be entered into the corresponding
moisture value box.
7.4.3
Manual Averaging When the Digital Input is Set to Average / Hold
It is possible to override the Automatic Averaging facility in the sensor so manual
averaging can take place. To temporarily disable the automatic averaging function in the
sensor touch the white box next to Averaging and select Manual (Figure 53). The sensor
will now disable the digital input and only start the averaging when manually started using
the Hydro-View. If the digital input is set to Average/Hold the averaging will go back to
automatic once the averaging screen is closed.
Summary of Contents for HV04
Page 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Page 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...