Step 3
To install a new radiator carefully take it out from the package - do not touch its transparent surface (in case
of any dirt - wash it with alcohol). Carefully slide the radiator into the housing, put on the joints, twist in the
grounding cord and turn on the device.
67
7.5 RO module replacement
The RO module lifetime is 3-4 years. The lifetime of membrane cartridges may vary due to the flow rate, its
characteristics and level/types of feed water contamination.
• Symptoms of incorrect membrane functioning:
• Lower system efficiency, slower flow rate,
• Shorter lifetime of ion exchange columns.
To replace an RO module open the back door of the demineralizer. To do it, twist off four screws on the sides
and take the door out. Disconnect the RO module from the tubes and replace it with a new one. Close the
door and screw the screws.
Reset:
Summary of Contents for HLP 10
Page 2: ......
Page 3: ...www hydrolab eu INSTRUKCJA OBSŁUGI DEMINERALIZATORA SERIA HLP PL ...
Page 5: ...5 ...
Page 7: ...HLP 10 HLP 20 HLP 30 HLP SMART HLP 5 7 ...
Page 9: ...9 300 200 300 200 500 443 232 232 300 200 HLP 5 HLP 10 30 HLP SMART ...
Page 38: ...38 ...
Page 39: ...www hydrolab eu USER MANUAL FOR HLP SERIE DEMINERALIZER EN 39 ...
Page 41: ...41 ...
Page 43: ...HLP 10 HLP 20 HLP 30 HLP SMART HLP 5 43 ...
Page 45: ...45 300 200 300 200 500 443 232 232 300 200 HLP 5 HLP 10 30 HLP SMART ...
Page 74: ...76 74 ...
Page 75: ...www hydrolab eu INSTRUCCIONES DE USO DEL DESMINERALIZADOR SERIE HLP ES 75 ...
Page 77: ...77 ...
Page 79: ...HLP 10 HLP 20 HLP 30 HLP SMART HLP 5 79 ...
Page 81: ...81 300 200 300 200 500 443 232 232 300 200 HLP 5 HLP 10 30 HLP SMART ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 115: ...115 HLP 10 HLP 20 HLP 30 HLP SMART HLP 5 ...
Page 117: ...117 300 200 300 200 500 443 232 232 300 200 HLP 5 HLP 10 30 HLP SMART ...
Page 146: ......