MCS 1000 Series
Measuring menu
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 53 / 76
BeWa MCS1000 3435427f V130 en-us 2012-06-15.doc
2012-06-15
CANCEL - cancel without saving
Parameters: none
SAVE - save changes
Parameters: none
Measuring menu
DSPLAY - Show measured variable
Function:
Select the measured variable that is displayed after
switching the unit on and after voltage has returned.
The measured variables FE A to NFE F show a
summation of the particles in the corresponding category.
The counter starts when voltage is switched on and is
reset by the absence of voltage. The counter readings are
stored in the internal memory after the passage of S.TIME
in each case.
The parameter CYC (cycle) gives the number of particles
per variable that were counted within the current
measurement time (parameter S.TIME).
The measuring time begins when the voltage is switched
on or the power-up menu is exited. The value of the
current interval is always displayed.
Parameters: FE
A:
ferromagnetic
particles (Fe)
Class A
FE
B:
ferromagnetic
particles (Fe)
Class B
FE
C:
ferromagnetic
particles (Fe)
Class C
NFE
D:
non-ferromagnetic
particles (nFe)
Class D
NFE
E:
non-ferromagnetic
particles (nFe)
Class E
NFE
F:
non-ferromagnetic
particles (nFe)
Class F
CYC
A:
ferromagnetic
particles (Fe)
Class A
CYC
B:
ferromagnetic
particles (Fe)
Class B
CYC
C:
ferromagnetic
particles (Fe)
Class C
CYC
D:
non-ferromagnetic
particles (nFe)
Class D