Technical data
Technical data
Measured value display
Continuous view in LCD display
Self-diagnosis
Continuous self-monitoring with error
indication in LCD display
Calibration
FCU 81xx: ISO 4402 / NAS 1638
Particle size channels:
FCU 81xx
:
2µm / 5µm / 15µm / 25µm / 50µm /
100µm
Measurement range (calibrated):
FCU 81xx
:
NAS 0 … 12
ISO 0 / 0 / 0 … 23 / 21 / 18
Indication range:
FCU 81xx
:
NAS 0 … 15
ISO 0 / 0 / 0 … 25 / 23 / 21
Recalibration
Recommended: annually
Battery-buffered log memory
Up to 3000 measured values
Connectors, rear:
INLET
Test point type 1620, DN2
OUTLET
Plug nipple, DN7
INLET
Operating pressure
1 … 250 bar
OUTLET
Flow rate
Maximum 250 ml/min at 250 bar
Operating pressure
max. 3 bar counterpressure
Measurement flow rate
40 … 60 ml/min
Permissible viscosity range
1 … 50 mm²/s
Fluid temperature range
0 … 70° C / 32 … 158° F
Voltage supply
24 V DC, ± 25%
FCU 8111 /-Aviation
en(us)
/
BEWA FCU8111 3690427a en-us 2017-01-23.doc
2017-01-23