HYDAC FILTER SYSTEMS
Operation |
BeWa FCU1315 4168086b
83 /
Fig. 42: Operating Mode - Measurement ≤ 45 bar
Make sure that the following sequence is observed:
✓
You have performed all of the steps listed in chapter
Putting the FCU into operation [
.
1. Check the pressure at the measurement location. The
existing pressure must be in the range of 1 … 45 bar /
14 … 650 psi.
WARNING! If the pressure is greater
than 45 bar / 650 psi, then use the high-
pressure adapter.
2.
Connect the entry hose (color: black) with the INLET
connection of the FCU (1).
Rotate the measurement coupling clockwise (2) to the
connection and tighten it by hand.
Summary of Contents for FCU1315
Page 3: ...1 HYDAC FILTER SYSTEMS 2 BeWa FCU1315 4168086b ...
Page 66: ...HYDAC FILTER SYSTEMS Operation BeWa FCU1315 4168086b 57 9 USB interface with cover ...
Page 76: ...7 HYDAC FILTER SYSTEMS 66 BeWa FCU1315 4168086b ...
Page 104: ...7 HYDAC FILTER SYSTEMS 94 BeWa FCU1315 4168086b Fig 46 BatteryPack in operation ...
Page 129: ...HYDAC FILTER SYSTEMS Performing maintenance BeWa FCU1315 4168086b 117 ...
Page 147: ...HYDAC FILTER SYSTEMS Finding rectifying errors BeWa FCU1315 4168086b 133 ...
Page 151: ...10 Finding rectifying HYDAC FILTER SYSTEMS 134 BeWa FCU1315 4168086b Error code 3 ...
Page 173: ...Table of HYDAC FILTER SYSTEMS 148 BeWa FCU1315 4168086b ...
Page 181: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 182: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 183: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 184: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 185: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 t t 1tFILTER SYSTEMS ...