Guidance and manufacturer’s declaration – electromagnetic immunity
The Hybresis Charging Station is intended for use in the electromagnetic environment specified below. The
customer or the user of the Hybresis Charging Station should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Hybresis Charging
Station, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
d
= [3,5]
√
P
V1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
d
= [3,5]
√
P
80 MHz to 800 MHz
E
1
d
= [7]
√
P
800 MHz to 2,5 GHz
E
1
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in met
(m).
ers
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Hybresis Charging Station is used exceeds the applicable RF compliance level above, the Hybresis Charging
Station should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Hybresis Charging Station.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
7