-------------------------------------------------------------------------------------SLIT LAMP HS-7500
31
10. Ordinary Maintenance
NOTE
10.1. Halogen Projection bulb
!
WARNING
!
CAUTION
The re pairs illustrate d be lo w sho uld all be c arrie d o ut with the po we r c able
disco nne cte d fro m the main supply plug. P le ase co ntac t the suppliers for any o ther
pro ble ms that c anno t be resolve d by the o pe ratio ns de s ibe d be lo w.
P le ase turn o ff the po we r switc h be fo re re plac ing lamps. P le ase place lamps after
c oo ling do wn. The he ate d lamp may burn yo ur finge rs.
Do n
’
t to uc h any o the r parts o f glass tube with the bare finge rs. Finge rprints, if any,
must be wipe d o ff fro m the glass tube with alc o ho l
To replace the Halogen projection bulb proceed as follow.
Before doing anything else, disconnect the plug from the main supply socket;
Remove the Lamp cover(39),
Remove the
(32) from the bulb stems;
Pull the burnt out bulb; be careful , it could be very hot;
Replace it with a new one(12V30W Halogen) using the plastic envelope
(never touch the bulb whit fingers)
l
l
l
l
l
Lamp socket