--------------------�SLIT LAMP HS-5000 3X 41
NOTE 1) These guidelines may not apply in all si
t
u
a
t
i
ons
.
Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
NOTE 2) lt is essential that the actual shielding effectiveness and filler
attenuation of the shielded location be verified to assure that they meet the
mínimum specification.
Field s trengths from fixed transmitters, such as base stations far radio
(cellular/cordless) telephones and land mobile radi
o
s
,
amateur radio, AM and FM
radio broadcast and TV broadcast cannot be p redicted theo retically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an
electromagnetic site survey should be oonsidered. lf the measured field strength
outside the shielded location in which the EUT is usad exceeds 3V/m, the EUT
should oe observad to verify normal operation.
lf abnormal pe rformance is observad, additional measures may be neces sary,
such as relocating the EUT or using a shielded location with a higher RF
shielding effectiveness and filler a
tt
e
nu
atio
n
.
41/46