Blocking Position Calibration
Procedures for blocking position calibration
①
Enter menu mode by pressing the
[MENU] button for 2 seconds.
②
Select “User Mode”.
③
Select “Calibration”.
④
Select “Blocking Position Calibration”.
⑤
Press the function([F1]~[F5]) button to
move the cross line.
⑥
The cross line must located in center of
the blocker.
\
!
NOTE
•
Place lens adaptor with LEAP tape to the lens adaptor setter and then turn the
switch lever to place the lens adaptor setter over the center of the LCD and the
push down the switch lever.
•
Through the small hole of the lens adaptor, you can easily adjust cross line to
the center position of the blocker.