Sonicaid Team Operator’s Manual
65
Checking analysis progress
After the first analysis has been performed at 10 minutes, you can check the key
results by pressing the return button on the keypad indicated by [RESULTS>>>] in
the display message bar. The Team shows the last calculated values for short term
variation, number of minutes of high variation and basal heart rate. An asterisk
beside a figure indicates an abnormal result. See Abnormalities, in Section 6.6.
For twins, the results shown are for the fetus on the currently selected audio
channel. This is confirmed on the display with either ULT-Y or ULT-B. If no audio
channel is selected, the default is the yellow channel.
To turn the analysis off
To record a trace without the analysis:
1 Press the printer button once to start the printer.
2 When Team asks you to enter the gestational age, select [ANALYSIS OFF].
3 The printer then starts without the analysis.
Or:
1 Press [MENU] once.
2 Select [TURN ANALYSIS OFF].
3 Press the printer button to start recording.
By default, the analysis is ON if it was on the last time Team was used, and OFF if it
was off the last time Team was used.
Analysing a stored trace
A stored trace can be analysed when it is printed on a Team Care printer. The patient
details for the stored trace must include the gestational age.
Summary of Contents for SonicaidTeam
Page 14: ...Sonicaid Team Operator s Manual 14 Addresses...
Page 68: ...Sonicaid Team Operator s Manual 68 Example printout of TeamCare trace and analysis report...
Page 69: ...Sonicaid Team Operator s Manual 69 Example printout of TeamCare twins analysis results...
Page 79: ...Sonicaid Team Operator s Manual 79 Printout showing Team Trend analysis results...