Functional Tests (Mechanical Units Only)
Tests must be carried out using a 220-240 V, 50 Hz power supply.
Test Setup and Initial Switch On
4.1
Set the pressure control knob fully counter-clockwise and ensure that both outlet ports are open
to atmosphere.
4.2
Connect the mains/power cable to a suitable power socket outlet and then set the on/off switch
to ‘on’ ( ). Verify that the green indicator in the switch is illuminated. Verify that the red
LOW
PRESSURE
indicator is illuminated. Use the fingers to close both outlet ports and verify that
the
LOW PRESSURE
indicator goes off.
Issue 2
Chap 5
Nov 2001
SER0004
Page 7
Mini Pump Service Manual
Testing