Maintenance
4-5
Fuse Changes
Refer to Chapter 5 for the proper fuse values.
To change the fuses in the support unit, perform the following steps.
1.
Turn off the power to the support unit and unplug it.
2.
Remove the slotted screws holding the support unit cover in place and open the cover.
3.
Open the fuse disconnect terminal blocks by pulling the pivotable fuse holder open into the locked
position with one finger.
4.
Open the side cover of the fuse holder to eject the fuse.
The fuse area
The motor controller fuses are
in positions 67 and 68 and the
power main fuses are in
positions 69 and 70.