12.2 GUIDANCE AND MANUFACTURER’S DECLARATION—EMISSIONS
The T7 is intended for use in Professional Healthcare Facility environment specified below. The customer or user of the T7
Cart should ensure that it is used in such environment.
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
CAUTION: Interference may occur in areas where the intensity of electromagnetic disturbances is high such as near active
HF surgical equipment and the RF shielded room of a medical electrical system for magnetic resonance imaging. Although
these environments are typically found in hospitals they are Special Environments and are not considered part of the
Professional Healthcare Facility Environment as defined in IEC 60601-1-2.
Emissions Test
Compliance Level
Electromagnetic
Environment—Guidance
RF Emissions CISPR 11
Group 1
The T7 uses RF energy only for its
internal functions. Therefore, its
RF
emissions are very low and are
not likely to cause any interference
with nearby electronic equipment.
RF Emissions CISPR 11
Class A
Harmonics IEC 61000-3-2
Class A
Flicker IEC 61000-3-3
N/A
120V AC 60 H
z input power.
The T7 is suitable for use in all
establishments, including
domestic, and those directly
connected to the public low
voltage power supply network
that supplies buildings used
for domestic purposes.
All Equipment and Systems
57