Secondary control panel
Indicator
The lift is equipped with a secondary control
panel located on the lift which can be used
as an alternative to the hand control. It is
equipped with Up button (2) and Down button
(3) for raising and lowering, an OFF/ON button
(1) for activation/deactivation of the lift and
an emergency lowering button (4) for quick
lowering the lift.
Press emergency lowering button and down
button at least 3 seconds simultaneously to
activate the emergency lowering. Older lifts
(before july 2016) may not have double key
command and delay.
The indicator light has the following indica
-
tions, which are only accessible when the lift is
switched on.
Flashing
green
The lift is on and ready for use. Batte
-
ry level is Full.
Flashing
yellow
The lift is on and ready for use. Batte
-
ry level is Normal. It might be a good
idea to charge up the lift.
Flashing
red
The lift is on and ready for use. Battery
level is Low. The lift must be charged
up immediately.
Constant
green
The lift is in operation or loaded.
Constant
red
The lift safety catch has engaged. (See
troubleshooting table)
Constant
yellow
The lift is charging. The lift’s function
is blocked.
Flashing
green /
yellow
The lift’s service interval has passed.
Service and annual inspection must be
performed immediately.
Flashing
red /
yellow
The lift’s service interval has passed.
Service and annual inspection must be
performed immediately. Battery level
is Low. The lift must be charged up
immediately.
1 2 3