background image

8

www.humancaregroup.com

The product label contains information about 

the device and is located on the right side, close 

to the mast.

This product has been tested by an accredited 

test institute and complies with the necessary 

requirements in accordance with medical 

requirements as follows:  EN 10535, EN 60601-1 

(Electrical Safety) and EN 60601-1-2 (EMC).  This 

product is CE marked.

Product Label

Symbols on the unit and product label:

CE Marking

Date of manufacture

Manufactured for Human Care

Maximum load

Instructions: Read the instructions and 

make sure you understand them before 

use
Part number/reference number

Serial number

Warning

The device contains electronics and it 

must be recycled appropriately
Applied part type B: approved for 

physical contact with patients (IEC 

60601-1)

Product Symbols

Expected Life

The expected lifespan of the Alzaro, in normal 

use and under normal circumstances, apart 

from slings, spreader bars and batteries, is 

10 years or 10,000 transfers when serviced 

according to the instructions.

Summary of Contents for 11000490

Page 1: ...Alzaro Part no 11000490 User Manual...

Page 2: ...3 Product Benefits 3 Warnings 4 General Care and Use 5 Components 6 Technical Specifications 7 Product Label 8 Product Symbols 8 Understanding the Components 9 Usage Warnings 13 Usage Guidelines 14 Pr...

Page 3: ...Safety Instructions WARNING symbol indicates a potentially hazardous situation which if not avoided could result in serious injury to the user or damage to property and or the device itself INFORMATIO...

Page 4: ...s exceeded Product is damaged in any way The sling is not securely and adequately fastened You have not read and understood all warnings and instructions Warnings Do not touch any parts which may have...

Page 5: ...ected by a professional with respect to the client s safety and needs Facility administrator and or risk management department should approve all patient handling protocols associated with this mobile...

Page 6: ...ancaregroup com 4 1 2 3 6 5 7 Components 6 Front Wheel 7 Rear Wheel with brake 8 Hand Control not shown 9 Spreader Bar not shown 1 Lift Arm 2 Push Handle 3 Actuator 4 Battery Control Box Unit 5 Leg Ad...

Page 7: ...reading legs IP Rating IPX4 Turning Radius 148cm 58 3 Wheels Front 10cm 4 Rear 12 5cm 5 with brake Battery 24V DC 2 x 12V 2 9Ah Li Ion accu 24V DC 2 9Ah Motor Lifting 24V DC 8 4A max 8kN 10 switch on...

Page 8: ...nit and product label CE Marking Date of manufacture Manufactured for Human Care Maximum load Instructions Read the instructions and make sure you understand them before use Part number reference numb...

Page 9: ...ox 1 Control functions up down 2 Emergency Stop 3 Mains socket Understanding the Hand Control Emergency Stop Use the emergency stop to turn off the lift when not in use to reduce the battery discharge...

Page 10: ...battery module to open Disconnect the battery connectors and remove the old batteries Insert the new batteries in the same direction as the old batteries with the connectors at the top and connect Not...

Page 11: ...th a very fast low noise and robust motor that is designed with a ball circulation spindle limit switches and automatic electric shut off in the event that the patient has come in contact with the arm...

Page 12: ...e brake is in the unlocked position line up the wheels with the chassis and the direction of travel Manoeuvering the Alzaro Do not lift the Alzaro arm higher than is necessary to move it If you want t...

Page 13: ...way Do not use if there are any signs of wear and damage on any part of the device Use product only on a level surface Do not use the lift on stairs escalators or an incline decline such as up or down...

Page 14: ...control box will interrupt and stop the lifting process To activate press the red button De activate the emergency stop by turning clockwise Attaching the Spreader Bar The Alzaro is delivered with a t...

Page 15: ...inue with lift If necessary advise patient that lift is commencing If using the self balancing crade attach sling clips to Alzaro lift arm clips Ensure all straps are straight and clips secure Using t...

Page 16: ...odium hypochlorite as recommended by the CDC 1 Weaker dilutions example 1 100 may be used but may not be in accordance with CDC recommendations Gloves and protective clothing should always be worn whe...

Page 17: ...ted Check the function on both the control box and the hand control if the control box works and the hand control does not then the hand control needs to be replaced The control box functions will not...

Page 18: ...ccording to cleaning instructions Weekly Maintenance Check the operation of the lift and the operation of the emergency stop Check that all parts move easily and without any noise Charge the batteries...

Page 19: ...visible damage or wear All components are present eg battery There is no wear on the cables wires There is no damage to the chassis Operation The emergency stop is functioning The control box is func...

Page 20: ...l info ca humancaregroup com HUMAN CARE UNITED STATES 8006 Cameron Street Suite K Austin TX 78754 USA Phone 512 476 7199 Fax 512 476 7190 Email info us humancaregroup com HUMAN CARE AUSTRALIA Unit 4 4...

Reviews: