Hukseflux VU01 User Manual Download Page 9

 

VU01 manual v1614  

 

 

 9/25 

3

 

Specifications of VU01 

3.1

 

Specifications of VU01 

VU01 is a high-quality ventilation unit that may be used with pyranometer models SR20, 
SR20-D2,  SR22,  SR25  and  SR25-D1,  and  pyrgeometer  models  IR20  and  IR20WS.  Its 
purpose  is  to  improve  the  dependability  of  the  measurement.  Measurement  accuracy 
improves  because  offsets  are  reduced.  Reliability  benefits  from  prevention  of  dew  and 
frost  formation  and  quick  evaporation  and  sublimation  of  water  and  snow.  For 
specifications  of  pyranometers  and  pyrgeometers,  please  consult  the  SR20,  SR20-D2, 
SR22, SR25, SR25-D1 and IR20 manuals.  
 

Table 3.1.1

 

Specifications of VU01 (continued on next page) 

 

VU01 SPECIFICATIONS 
  

Compatible instruments 

SR20, SR20-D2, SR22, SR25, SR25-D1, IR20, IR20WS 

Compliance with standards 

ISO/TR 9901 

 

BSRN operations manual 

Rated operating power 

7.8 W at 12 VDC (unheated) 

Rated operating voltage range 

10.8 to 13.2 V 

Rated air flow 

50 m

3

/hr 

Rated operating temperature range 

-40 to +70 

o

Fan status signal 

2.5 VDC (optional use) 
High = fan OK = 2.5 V 
Low  = fan NOT OK = 

 0.4 V 

Heating 

5 and 10 W at 12 VDC (optional use) 

Increase of air temperature 

0.5 

o

C at 0 W heating; 1.0 

o

C at 5 W heating;  

1.5 

o

C at 10 W heating 

Zero offset a SR20 ventilated 

2.5 W/m

2  

at 0 W heating 

Offset SR20 by heating 

-2 W/m

2  

at 5 W heating 

Offset SR20 by heating 

-4 W/m

2  

at 10 W heating 

Offset IR20 by heating 

0 W/m

2  

at 5 W heating 

Offset IR20 by heating 

0 W/m

2  

at 10 W heating 

Cable resistance 

0.3 

Ω/m (2

 

x 0.15 Ω/m)

 

Voltage drop at 7.8 W ventilation 

0.2 V/m 

Voltage drop at 10 W heating power 

0.12 V/m 

Footprint diameter 

0.17 m 

Fan rotation rate 

5050 to 5400 1/min at 12 VDC  

Standard cable length (see options) 

5 m 

Cable diameter 

5.3 x 10

-3

 m 

VU01 connector 

M16 straight connector, male thread, 10-pole 

VU01 connector type 
 

HUMMEL AG 7.820.400.000 straight connector, male 
thread, for cable 5 to 9 x 10

-3

 m 

Cable connector 

M16 straight connector, female thread, 10-pole 

Cable connector type 

HUMMEL AG 7.810.300.00M straight connector, female 
thread, for cable 3 to 6 x 10

-3

 m, special version 

Connector protection class 

IP67 / IP69K per EN 60 529 (connected) 

Cable replacement 

replacement cable with cable connector can be ordered 
separately from Hukseflux 

IP protection class VU01 

IP54  

Summary of Contents for VU01

Page 1: ...USER MANUAL VU01 Ventilation unit for SR20 and SR22 pyranometers and IR20 pyrgeometers Copyright by Hukseflux manual v1620 www huksefluxusa com info huksefluxusa com...

Page 2: ...he VU01 cable and sensor cable will not damage the sensor but the instrument and ventilation unit will not function In case of cable replacement or extension beyond 5 metres by the user please note th...

Page 3: ...stallation of VU01 12 5 1 Installation of the ventilation unit 12 5 2 Installation of the sensor 12 5 3 Installation of the VU01 cover 13 5 4 Electrical connection of VU01 14 5 5 Power supply and cabl...

Page 4: ...in seconds t s Rotation rate N 1 min Supply voltage Usupply V Voltage drop over cable Udrop V Fan voltage supply Ufan V Cable resistance Rcable Wire resistance per meter Rewire m Cable length Lcable...

Page 5: ...and SR25 D1 secondary standard pyranometers and IR20 and IR20WS research grade pyrgeometers Ventilation promotes thermal equilibrium between all components of radiometers and thereby reduces zero off...

Page 6: ...tor power is relatively high in order to increase the airflow The heaters should be activated using external relays A fan status signal high low may be used by the user to monitor rotation of the fan...

Page 7: ...are 2 x M5x60 screw 2 x M5 retaining torque nut 4 x 5 x 10 3 m washer spare filter 1 x product certificate any other options as ordered 1 3 Options longer cable in multiples of 5 metres Specify total...

Page 8: ...tem below cover 9 VU01 connector 10 VU01 cable and cable connector Ventilation promotes thermal equilibrium between all components of radiometers and thereby reduces zero offsets Dew and frost formati...

Page 9: ...fan OK 2 5 V Low fan NOT OK 0 4 V Heating 5 and 10 W at 12 VDC optional use Increase of air temperature 0 5 o C at 0 W heating 1 0 o C at 5 W heating 1 5 o C at 10 W heating Zero offset a SR20 ventil...

Page 10: ...cluding 5 m cable 1 35 kg Packaging box of 330 x 250 x 220 mm VERSIONS OPTIONS ACCESSORIES Longer cable in multiples of 5 metres option code total cable length consult Hukseflux see chapter 5 5 on ins...

Page 11: ...ultra high accuracy measurements the following manual may serve as a reference Baseline Surface Radiation Network BSRN Operations Manual Version 2 1 L J B McArthur April 2005 WCRP 121 WMO TD No 1274...

Page 12: ...nsor mounting plate to the bottom of the sensor using two M5x10 screws See Figure 5 2 1 A T25 torx key to do so is included with VU01 Affix the sensor with the sensor mounting plate to the ventilation...

Page 13: ...ver using thumbscrew 1 Connect the VU01 cable connector to the VU01 connector 4 To remove cover Loosen thumbscrew 1 by turning counterclockwise until the tapered part of the thumbscrew is ejected from...

Page 14: ...and a power supply as short as possible For cable extension see the next paragraph and appendix on this subject Table 5 4 1 The electrical connection of VU01 The heaters and fan status signal are not...

Page 15: ...e ventilator a high cable resistance casuses a high voltage drop VU01 power consumption is 7 8 W at a low voltage 12 VDC The current that is needed 0 66 A causes a significant voltage drop Using a 12...

Page 16: ...V V V V V 1 0 2 0 1 0 1 0 0 0 0 0 0 0 0 5 1 0 5 0 3 0 2 0 1 0 1 0 1 10 1 1 0 7 0 4 0 3 0 2 0 1 15 1 0 0 6 0 4 0 3 0 2 20 0 9 0 5 0 3 0 2 30 0 8 0 5 0 3 40 0 7 0 4 50 0 9 0 5 Examples For a cable leng...

Page 17: ...per minute The signal is 2 5 V When the fan is running too slowly or not at all a low signal US low is provided The fan status signal will only function when VU01 itself is powered Figure 5 6 1 A fan...

Page 18: ...signal between the blue fan status signal and yellow fan status signal wire as shown in Figure 5 6 2 Figure 5 6 2 Wiring diagram required to use VU01 s fan status signal To use the optional status sig...

Page 19: ...shooting Table 6 3 1 Trouble shooting for VU01 PROBLEM SOLUTION Fan and or heaters do not function check wiring for correct connection of and check power supply Fan does not work check for obstructio...

Page 20: ...VU01 manual v1614 20 25...

Page 21: ...ront side the filter retainer Turn the screwdriver 90 to detach the filter retainer from the finger guard Replace the used filter cloth with a fresh filter spare part VU01F Slide the filter retainer u...

Page 22: ...rred specifications for VU01 cable replacement and extension General replacement please order a new cable with connector at Hukseflux or choose a DIY approach In case of DIY replacement by the user se...

Page 23: ...quirements of the following directive 2004 108 EC The Electromagnetic Compatibility Directive hereby declare under our sole responsibility that Product model VU01 Product type Ventilation unit has bee...

Page 24: ......

Page 25: ...2016 Hukseflux Thermal Sensors B V www hukseflux com Hukseflux Thermal Sensors B V reserves the right to change specifications without notice...

Reviews: