27 | 60
OWNERS MANUAL | Cobolt 08-01 Series | D0141-K SEPTEMBER 2020
Direct On/Off control
06-MLD
This feature is not available for 06-MLD.
06-DPL
The Direct On/Off control feature enables turning the laser On/Off using a 5 VDC signal. After having configured the
laser for Direct Input operation (factory set or by executing @cobasdr 1), the laser can only start-up when 5 VDC (max
12.5 VDC) is applied to pin 3 on the Molex connector with 0 VDC on pin 2 as reference. Shifting the signal to
0
VDC on
pin 3 will turn the laser off and put the laser in stand-by mode. This input only controls the on/off state of the laser and
cannot be used to modulate the power output. The remote interlock connection between pin 1 and 2 must also be
made as described in section 5.8 : Remote Interlock Connector.
Molex connector on back side of laser head.
NOTICE
This function is not available for CDRH compliant models and cannot be used with key control
box.