![Huber CC-405 Operating Manual Download Page 61](http://html1.mh-extra.com/html/huber/cc-405/cc-405_operating-manual_2180789061.webp)
61
Operation as bath thermostat
(valid for temperature control units with
baths)
Please take into account the change of volume that may occur when placing samples
(e.g. an Erlenmeyer flask) into the bath. Place the sample into the empty bath. Only
then start filling the bath with sufficient thermal fluid. Note that the fluid level will sink
when taking the sample out of the bath. This may lead to a later safety switch off (low
level protection) during temperature control. Always switch off temperature control
before taking the sample out of the bath.
e.g. K12-cc-NR Variostat with Variostat bath
Please note, that combinations using the Variostat with Variostat baths do not have the
whole space (and volume) of the Variostat bath available (see picture above). For bath
thermostats note the bath opening which is given in the catalogue and data sheet.