F30 Set date
Setting the date.
F31 Set time
Setting the time.
F33 Set over-temperature protection
Setting the over-temperature protection. Please note chapter on setting the
over-
temperature switch.
F34 Air purge
Start / Stop air purge
F35 Compressor automatic
This is used to select the operation of the compressor: Default setting is
always ON
Automatic
:
The compressor automatic is set to switch on and off as required by the unit.
Benefit: Energy saving
Disadvantage: Longer response times to sudden increase in cooling demand.
Always ON:
The compressor is always running, so the refrigeration system is always immediately
available.
Always OFF:
The compressor is always off.
The compressor automatic (valid for units with compressors only) must be set to
always on
when selecting
process security
in the main menu
Over-temperature protection / OT Mode
F37 Temperature mode
Setting the temperature mode.
F39 Signal
Activating the signal.
F40 Test RS232
Start / Stop
F41 Select user menu
Please refer to chapter
select user menu
.
F42 Configure user menu
Please refer to chapter
configure user menu.
F46 Define analogue input (function analogue input)
This function allows an analogue input current of 0/4-20 mA or 0-10V to be assigned
to an input value.