3

Perfect protection function, and the spare fuse prompt functions.

III. Specification Parameter

Safety Standard

IEC 61010-1

CAT.IV 600V

IEC 61326

EMC specification: test, control and electrical equipment for testing

IEC60529 IP64

Outer closed state

● AC Power

220V±10%

50 HZ

20 VA

● battery

16.8V Lithium ion rechargeable battery

● Battery Life

5000V@100M

6 hour

● size

L x W x H

):

26cm x 20cm x 16cm

● weight

3kg

● Test voltage accuracy: the nominal value 100% to 110%

● Voltage measurement accuracy

5%+3V

● Current test range

10mA

● current measure precision

5%+0.2nA

● short-circuit current

5mA

● Capacitance test range

20uF

● Capacitance testing accuracy

15%+0.03uF

● Capacitor discharge rate

from 5000V to 10V

0.5S/µF

● Insulation resistance testing range and accuracy

T:23±5ºC

RH :45 – 75%

range

accuracy

500V

1000V

2500V

5000V

Unspecified

<100k

<100k

<100k

<100k

5%

100k-10G

100k-20G

100k-100G

100k-100G

20%

10G -100G

20G-200G

100G-1T

100G-1T

Summary of Contents for HZJY-5K-I

Page 1: ...HZJY 5K I Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...Contents I Safe Use Items 1 II Feature 2 III Specification Parameter 3 IV Introduction Of Instrument 4 V Instrument Using 7 VI Packing List 8...

Page 4: ...tions If you don t comply measurement may cause personal injury and equipment damage Dangerous for improper operation can lead to serious or fatal damage Before use please wear insulated gloves Please...

Page 5: ...ltage insulation resistance tester with 5 range 500V 1000V 2500V 5000V test up to 5T Strict accordance with the safety standards for design Insulation resistance range 5T 5 kv Short circuit current ca...

Page 6: ...our size L x W x H 26cm x 20cm x 16cm weight 3kg Test voltage accuracy the nominal value 100 to 110 Voltage measurement accuracy 5 3V Current test range 10mA current measure precision 5 0 2nA short ci...

Page 7: ...g 2 AC ON 3 Test Port E G L 4 LCD display 5 Key Key and function Key numbered from left to right the first for a total of five off down the power Key key number Description 1 Press the switch on butto...

Page 8: ...pass filter applicable to most situations 2 software Low pass filter High resistance greater than 100G recommended choose 3 hardware software low pass filter high resistance is greater than 1T recomm...

Page 9: ...OK Confirm to enter the current menu function or to determine the current Settings and exit Help menu system HELP Record the serial number of product production date version information Insulation tes...

Page 10: ...e polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the...

Page 11: ...8 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Red test line 1 4 Black test line 1 5 Green test line 1...

Reviews: