V. Instrument Using

measured DAR

:

1. Select more than 1 minute length measurement.

2. Select the appropriate voltage and start.

3. The instrument automatically record 15 s and 60 s resistance, and calculate the

absorption than after 60 seconds.

Absorption ratio calculation method: DAR = R60 / R15;

measured PI :

1. Select more than 10 minute length measurement.

2. Select the appropriate voltage and start.

3. The instrument automatically record 15S, 60S and 600S resistance, and calculate the

polarization index after 600 seconds.

Polarization index calculation method : PI= R600/R60;

test connection points for attention

1.

Tested to confirm product safety grounding, try not charged.

2.

Confirm the instrument E end (ground) have been grounded.

3.

The use of G terminal (protection ring) (this machine is of low voltage side of the

screen)

When measuring high insulation resistance, the sample should be in the position a

conductor surface protection between the two measuring end ring, and the conductor

protection ring is connected to the instrument with a test line G side, try to eliminate the

measuring error caused by the surface leakage current, security testing is accurate.

Summary of Contents for HZJY-15K

Page 1: ...HZJY 15K Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...Contents I Warning 1 II Feature 2 III Specification Parameter 2 IV Introduction of Instrument 3 V Instrument Using 7 VI Packing List 8...

Page 3: ...erators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in ele...

Page 4: ...eaches 2 mA still ensure accuracy of test instruments 2 methods for power use of lithium ion batteries battery life can be up to 4 hours 10000V 100M Insulation resistance simulation column Digital fil...

Page 5: ...tance testing range and accuracy T 23 5 C RH 45 75 range accuracy 500V 1000V 2500V 5000V 10000V Unspecified 1M 1M 1M 1M 1M 5 1M 10G 1M 20G 1M 100G 1M 100G 1M 200G 20 10G 100G 20G 200G 100G 1T 100G 1T...

Page 6: ...iption OFF Power Off No operation 3 minutes automatic shutdown H V Set test voltage Manual 500V 1000V 2500V 5000V 10000V 15000V TIME SetThe test time 01 00 10 00 30 00 Switching software keys to a new...

Page 7: ...p 1 In shutdown state hold key 2 the 2 key 2 Press the 1 key the power button until into the contrast adjustment menu 3 Press 1 or 2 keys to adjust contrast until the right 4 Press the 4 key save and...

Page 8: ...Regulating key Soft key Description Regulating key By a voltage rise of 10V long time boost up the fastest up to 600V per second Regulating key According to a voltage drop 10V long time step down spe...

Page 9: ...e polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the...

Page 10: ...8 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Red test line 1 4 Black test line 1 5 Green test line 1...

Reviews: