Squared before measurement, confirm the voltage selection in the appropriate value.

If long time not to use, the battery should be fully charge, and recharge in 3 months.

Squared do not in high temperature, damp, places and condensation may be placed

under direct sunlight for a long time.

Please use a damp cloth or cleaner to clean the equipment enclosure, do not use

abrasives or solvents.

Instrument wet, please dry storage first.

II.Features

It has 5 ranges: 500V, 1000V, 2500V, 5000V, 10000V, the maximum test can reach 20TΩ,

and the manual boost function.

● Designed in strict accordance with safety standards

● Insulation resistance range 20TΩ@10kV

● The short-circuit current can be adjusted up to 3mA.

● Automatically display the test values ​ ​ of polarization index (PI) and absorption ratio

(DAR), which can test leakage current and capacitance, as well as dielectric discharge

test (DD), step voltage test (SV), ramp test: the instrument automatically boosts, Until the

set value, it is qualified. If it is not reached, there is a breakdown in the middle or the

current is too large, it is unqualified.

● Excellent anti-interference performance, when the interference current reaches 2mA,

the instrument still guarantees the test accuracy.

● The AC and DC voltage test function of the tested circuit can automatically identify AC or

DC.

● The capacitive test product is discharged quickly. When the cable is tested, manual

discharge is not required, and the instrument automatically discharges quickly.

● 2 power modes: use lithium battery for power supply, battery life can reach 6 hours.

At the same time, it can be charged in use. In the event of a power failure, it can

automatically switch from AC power supply to battery power supply.

● All Chinese display, Chinese prompt operation, simple and clear, and can display the

insulation resistance simulation column.

Summary of Contents for HZJY-10K-IA

Page 1: ...HZJY 10K IA Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...al and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this inst...

Page 3: ...Contents I Safe Use Items 1 II Features 2 III Technical specifications 3 IV Instrument introduction 4 V Instrument use 5 VI Appendix 7...

Page 4: ...erious or fatal damage Before use please wear insulated gloves Please do not over AC DC600V measurement circuit Please do not test in flammable places spark could cause an explosion please do not hand...

Page 5: ...nt and capacitance as well as dielectric discharge test DD step voltage test SV ramp test the instrument automatically boosts Until the set value it is qualified If it is not reached there is a breakd...

Page 6: ...l value Output voltage monitoring accuracy 5 10V Voltage measurement range AC 30 600V 50HZ 60HZ DC 30 600V Voltage measurement accuracy 2 3dgt Current test range 10mA Current measurement accuracy 5 0...

Page 7: ...utton 6 High voltage indicator Keys and function description There are 5 buttons in total from left to right The first one is the power on button in the shutdown state Button No Instruction 1 Power on...

Page 8: ...When charging the power indicator light is on and the battery indicator on the display screen displays from 0 grid to full grid When it is full it displays full grid System help menu description Help...

Page 9: ...stance values of 15S 60S and 600S and calculates the polarization index after 600 seconds Polarization index calculation method PI R600 R60 DD test This test is suitable for the diagnosis of multilaye...

Page 10: ...should be connected to the G terminal of the meter with a test wire to eliminate the leakage current caused by the surface of the test product The measurement error ensures the accuracy of the test Es...