background image

1

I. Overview

Many power systems, railway systems, large-scale petrochemical plants and

enterprises have a lot of electrical equipment whose internal insulation are mostly oil-filled

insulation type, and therefore, test on insulating oil dielectric strength is common and

necessary. To meet the needs of the market, we have developed and produced a series of

insulating oil dielectric strength testers according to national standard GB/T507-2002,

industry standard DL429.9-91 and the latest Electric Power Industry Standard

DL/T846.7-2004 by ourselves. This instrument, by using a single-chip microcomputer as

the core, can operate in full automation with high accuracy, greatly improving work

efficiency and reducing the labor intensity of workers. Moreover, it is small in size and

convenient to carry.

II. Key Function And Feature

1. With a microprocessor, automatically fulfill the withstand voltage test for oil circulation

with a range of 0

80KV (including boosting, maintaining, mixing, standing,

calculation, printing and other operations)

2. Display by large-screen LCD and prompt by English menu.

3. The instrument operation is simple. The machine will automatically complete the

withstand voltage test on one cup of sample oil after simple setting by the operator.

Breakdown voltage value of 1 to 6 times and recurrent times will be automatically

saved. After the test, the thermal printer will print each breakdown voltage value and

average value.

4. Maintain power down. It can save 100 tested results and displaying current ambient

temperature and humidity.

5. Adopt a single-chip microcomputer to boost the voltage at an even speed. The

voltage is accurate at 50HZ, ensuring the whole process easy for control.

6. With overvoltage, over current, limiting protection, in order to ensure the safety of

operators.

7. With a temperature measurement display function and system clock display.

8. Communicate with computer with a standard RS232 interface.

Optional

Summary of Contents for HZJQ-N1

Page 1: ...HZJQ N1 Transformer Oil BDV Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrum...

Page 3: ...I Key Function And Feature 1 III Major Technical Indicators 2 IV Panel Description 2 V Operational Approach 3 VI Precautions 8 VII Maintenance 8 VIII Oil Cup Cleaning Method And Common Fault Clearance...

Page 4: ...l the withstand voltage test for oil circulation with a range of 0 80KV including boosting maintaining mixing standing calculation printing and other operations 2 Display by large screen LCD and promp...

Page 5: ...uency 1 6 time s optional 7 Booster capacity 1 5KVA 8 Measuring accuracy 3 9 Supply voltage AC220V 10 50Hz 1 Hz 10 Power 200W 11 Applicable temperature 0 45 12 Applicable humidity 75 RH 13 Overall dim...

Page 6: ...ide of the equipment to the earth wire firmly before start the equipment Sample the oil according to relevant standard Adjust the electrode distance inside the oil cup with standard gauge Clean the cu...

Page 7: ...ng time 0 15 min Pausing interval 0 5 min Stirring time 10 S Boosting times 1 6 Boosting speed 2 0 3 5KV S Automatic printing no 08 08 14 16 36 21 0 5 5KV S Setting of voltage boosting the user may se...

Page 8: ...10s 20s 30s continuous Boosting times 1 6 times Boosting speed 0 5KV S 5KV S Press Back key to exit this interface after the setting is done Start Experiment Press the Select key to select the Start...

Page 9: ...lue 00 0KV 1 78 9KV Start Stop Step up Step down Back 27 55 RH 08 08 14 16 30 44 Start Stir Step down Delay Back 288 Ending the first test continue to the next until completed the boost times of the u...

Page 10: ...the Confirm key to enter the following interface 27 55 RH 08 08 14 16 32 41 Average Value 72 8KV 1 78 9kV 2 75 8kV 3 73 1kV 4 77 9kV 5 72 5kV 6 58 9kV Standard Deviation 6 6kV Coefficient of variatio...

Page 11: ...during the operation VII Maintenance 1 This equipment shall not be exposed in moist environment 2 Keep the oil cup and the electrodes clean Fill the cup with fresh transformer oil for protection duri...

Page 12: ...ng Blade Cleaning Method Wipe the agitating blade again and again with clean silk cloth until fine particles are not found on their surfaces It is forbidden to touch the surfaces with hands Use forcep...

Page 13: ...Check break of high voltage line Light display contrast Adjust the potentiometer on the circuit board Printer failure Check plugging of printer power line Check plugging of printer data line IX Packin...

Reviews: