Use petroleum ether (other organic solvents are forbidden) to clean three times. Each

time shall follow the bellow procedures:

Pour the petroleum ether into the oil cup till the cup is 1/4

1/3 full.

Cover the cup mouth with a piece of glass cleaned by petroleum ether. Shake the

cup evenly for one minute with certain force.

Pour away the petroleum ether and dry the cup with a blower for 2~3 minutes.

Use the oil sample to be tested to clean the cup for 1~3 times.

Pour the oil sample to be tested into the oil cup, about 1/4

1/3 .

Cover the oil cup mouth with a piece of glass cleaned . Shake the cup evenly for

one to two minute with certain force.

Pour away the left oil sample and then can do the suppress experiment.

2.

Agitating Blade Cleaning Method

Wipe the agitating blade again and again with clean silk cloth until fine particles are not

found on their surfaces. It is forbidden to touch the surfaces with hands.

Use forceps to clamp the blade; put them into petroleum ether and wash.

Use forceps to clamp the blade and dry them with a blower.

Use forceps to clamp the blade; put them into the oil sample to be tested and wash.

3

Oil Cup Storage

Method 1

Fill the cup with good insulating oil after the test is finished and place it stable.

Method 2

Clean and dry the cup under the above procedures and then put it into a

vacuum dryer.

Note: The oil cup and agitating blade shall be cleaned under the above procedures after

the first test and tests with poor oil.

4

Common Fault Clearances

Power light off, screen display off

Check the plugging of power plug.

Check the condition of the protector tube inside the power plug.

Check the socket electricity.

Summary of Contents for HZJQ-N1

Page 1: ...HZJQ N1 Transformer Oil BDV Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrum...

Page 3: ...I Key Function And Feature 1 III Major Technical Indicators 2 IV Panel Description 2 V Operational Approach 3 VI Precautions 8 VII Maintenance 8 VIII Oil Cup Cleaning Method And Common Fault Clearance...

Page 4: ...l the withstand voltage test for oil circulation with a range of 0 80KV including boosting maintaining mixing standing calculation printing and other operations 2 Display by large screen LCD and promp...

Page 5: ...uency 1 6 time s optional 7 Booster capacity 1 5KVA 8 Measuring accuracy 3 9 Supply voltage AC220V 10 50Hz 1 Hz 10 Power 200W 11 Applicable temperature 0 45 12 Applicable humidity 75 RH 13 Overall dim...

Page 6: ...ide of the equipment to the earth wire firmly before start the equipment Sample the oil according to relevant standard Adjust the electrode distance inside the oil cup with standard gauge Clean the cu...

Page 7: ...ng time 0 15 min Pausing interval 0 5 min Stirring time 10 S Boosting times 1 6 Boosting speed 2 0 3 5KV S Automatic printing no 08 08 14 16 36 21 0 5 5KV S Setting of voltage boosting the user may se...

Page 8: ...10s 20s 30s continuous Boosting times 1 6 times Boosting speed 0 5KV S 5KV S Press Back key to exit this interface after the setting is done Start Experiment Press the Select key to select the Start...

Page 9: ...lue 00 0KV 1 78 9KV Start Stop Step up Step down Back 27 55 RH 08 08 14 16 30 44 Start Stir Step down Delay Back 288 Ending the first test continue to the next until completed the boost times of the u...

Page 10: ...the Confirm key to enter the following interface 27 55 RH 08 08 14 16 32 41 Average Value 72 8KV 1 78 9kV 2 75 8kV 3 73 1kV 4 77 9kV 5 72 5kV 6 58 9kV Standard Deviation 6 6kV Coefficient of variatio...

Page 11: ...during the operation VII Maintenance 1 This equipment shall not be exposed in moist environment 2 Keep the oil cup and the electrodes clean Fill the cup with fresh transformer oil for protection duri...

Page 12: ...ng Blade Cleaning Method Wipe the agitating blade again and again with clean silk cloth until fine particles are not found on their surfaces It is forbidden to touch the surfaces with hands Use forcep...

Page 13: ...Check break of high voltage line Light display contrast Adjust the potentiometer on the circuit board Printer failure Check plugging of printer power line Check plugging of printer data line IX Packin...

Reviews: