4

Figure 4

The ">,<" symbols appear on the interface to change alternately, indicating that it is in the

measurement state and the measurement can be performed. After a set of data is

measured, press the measurement button, and the internal air pump will stop. If you need

to measure the next data, press Measure button, the air pump inside the machine starts

and starts measuring again, and it can cycle like this. (Attention!!!!!Do not turn off the

power on key during the entire measurement cycle, as this will require reheating for 90

seconds before the measurement can be performed).

The entire measurement time can be set at the initial time, and the measurement will be

automatically stopped when the time comes.

During the measurement process, according to the concentration alarm limit set by the

user, the screen will appear: SF6 gas high-risk concentration alarm voice prompt, and the

buzzer will sound the alarm.

After the measurement is completed, you need to save the measured data and press the

save button (note! This save button only saves the data to the memory, It will be lost when

shut down, it needs to be saved permanently. Enter the menu map and select permanent

save. The data will be automatically saved after the next startup Read into memory, data

can save up to 8000 groups.

(2) Touch key operation

After powering on, after the interface shown in Figure 2 appears, touch anywhere on the

screen, the interface shown in Figure 5 appears:

Summary of Contents for HZIR-1000

Page 1: ...HZIR 1000 SF6 infrared quantitative detector User Manual Huazheng Electric Manufacturing Baoding Co Ltd ...

Page 2: ...anual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrument If you are satisfied with this instrument please tell your colleagues if you are not satisfied with this instrument please call 0312 6775656 to tell you to serve you at all times Baoding Hua...

Page 3: ...Contents I Overview 1 II MainTechnical index 1 III Main functions and features 2 IV Method of operation 2 V Matters needing attention 7 ...

Page 4: ...le and high precision performance II MainTechnical index 1 Technical parameters 1 Minimum detection value 0 1ppm 2 Detection range 0 1000ppm 3 Response time 3 seconds 4 Recovery time 5 seconds 5 Reference error 1 0 1000 ppm 5 1000 5000 ppm 6 Repeat error 2 7 Display mode 3 5 inch color touch screen display 8 The detection length of the probe gun straight rod type 130mm 9 Continuous working time 8 ...

Page 5: ... exported through USB interface 7 Four cell lithium polymer battery reliable and explosion proof 8 The first temperature control system for the whole machine 9 Advanced power management system automatic standby and shutdown 10 Compact structure beautiful appearance concise content and simple operation IV Method of operation 1 Turn on Press the power button the display is on and the interface shown...

Page 6: ...hen the interface shown in Figure 2 appears after power on press the measurement key and the interface shown in Figure 3 appears Figure 3 Automatically delay preheating for 90 seconds countdown the sensor itself requires It must be preheated to enter a stable working state When the timer is 0 seconds the built in air pump is working the left indicator light the interface shown in Figure 4 appears ...

Page 7: ...he initial time and the measurement will be automatically stopped when the time comes During the measurement process according to the concentration alarm limit set by the user the screen will appear SF6 gas high risk concentration alarm voice prompt and the buzzer will sound the alarm After the measurement is completed you need to save the measured data and press the save button note This save but...

Page 8: ...gure 2 will be restored after touching the screen Touch the measurement key measurement can be performed The operation procedure is the same as the shortcut key measurement procedure Touch the setting key the interface shown in Figure 6 appears Figure 6 Users can make initial settings according to their needs and the setting content is as described above Touch the display key several interfaces in...

Page 9: ... and any part of the screen can restore the backlight In the measurement state the backlight will not be completely turned off only the brightness will be reduced When the help button is touched on the screen a simple instruction for using the instrument appears for the user to read 2 Shut down After the measurement is over press the power switch key the display does not display the touch screen d...

Page 10: ...ical instructions and instructions in the manual 7 The operator must be familiar with the safety measures safe operation and cleaning and maintenance of this equipment 8 Engineers and technicians with certain professional knowledge are required to operate this equipment 9 The operator must read this manual thoroughly and understand the importance of SF6 gas measurement safety 10 The personnel mana...

Reviews: